A flame atomic absorption spectrometric method is described for the determination of platinum, palladium, and rhodium in used and fresh alumina-supported and sieve catalysts. A lanthanum solution was used to eliminate chemical matrix interferences. Samples were completely dissolved in sulfuric acid and aqua regia. Overall, the method has resulted in % RSD values ranging from 1.1-5.7, 2.2-5.9, and 4.8-5.3 for Pt, Pd, and Rh, respectively; the 3 sigma detection limits were 5 mu g/L Pt, 3 mu g/L Pd, and 40 mu g/L Rh.
机译:描述了一种火焰原子吸收光谱法,用于测定用过的和新鲜的氧化铝负载的和筛分的催化剂中的铂,钯和铑。镧溶液用于消除化学基质干扰。样品完全溶于硫酸和王水。总体而言,该方法得出的Pt,Pd和Rh的%RSD值分别为1.1-5.7、2.2-5.9和4.8-5.3。 3 sigma的检出限为5μg / L Pt,3μg / L Pd和40μg / L Rh。