A three-valve, pulsatile tubular pump was used in 24 pigs weighing 10.2 +/- 3.2 kg; the pump was connected to a neonatal hemofiltration circuit. Ninety-two periods of 30 min were studied to analyze the efficacy of the system with variations in the percentage time in diastole, the diastolic speed, the systolic speed, and the percentage time in systole during which the postfilter valve was closed. System efficacy was determined by the blood flow through the filter, the ultrafiltrate volume, the vascular overload measured by the inlet aspiration pressure, and the filter overload measured by the cross-filter pressure drop and the transmembrane pressure. The variations in the pump parameters did not lead to significant differences in the efficacy of the system or in the vascular or filter overload. The parameters must be adjusted in each case to obtain the best yield with the lowest vascular and filter overload.
机译:在24头体重为10.2 +/- 3.2 kg的猪中使用了三阀脉动管状泵。泵连接到新生儿血液过滤回路。研究了九十二个30分钟的时间,以分析系统的功效,其中舒张压的百分比时间,舒张速度,收缩速度和收缩压百分比时间(后过滤器阀关闭)的变化。系统功效取决于通过过滤器的血流量,超滤液量,通过入口抽吸压力测量的血管超负荷以及通过交叉过滤器的压降和跨膜压力测量的过滤器超负荷。泵参数的变化并未导致系统功效或血管或过滤器超负荷的显着差异。必须在每种情况下调整参数,以在最低的血管和过滤器过载情况下获得最佳产量。