Cell salvage devices are routinely used to process red blood cells (RBCs) shed during cardiac surgery. The purpose of this study was to evaluate three commercially available cell saver (CS) devices in terms of erythrocyte function and the quality of washed RBCs during cardiopulmonary bypass (CPB). Thirty patients undergoing CPB were randomly allocated to three CS devices: Group C (Cell Saver 5+; Haemonetics, n=10), Group M (autolog; Medtronic, n=10), and Group F (CATS; Fresenius HemoCare, n=10). Blood samples were collected from reservoirs and transfusion bags. Reservoirs and washed RBCs were analyzed for erythrocyte aggregation index, deformation index (DI) and hematocrit viscosity, 2,3-diphosphoglycerate (2,3-DPG), hematocrit (Hct), hemoglobin (Hb), free Hb removal (ΔfHb), glucose (Glu), lactate (Lac), and blood urea nitrogen. After processing, Groups C (P=0.026) and M (P=0.032) had relatively higher erythrocyte DI compared with Group F. Group C had lower Δ2,3-DPG compared with Groups M (P=0.001) and F (P=0.001). Group F provided the maximum concentration of Hct (P=0.021; 0.046) and Hb (P=0.008; 0.013). In addition, Groups C (P=0.035) and M (P=0.038) had a higher removal of fHb (ΔfHb), differing significantly with Group F. In conclusion, CS devices use the same theory of centrifugation; however, based on different designs, the function of the washed erythrocyte and undesirable content removal efficiency differs widely from one device to another.
机译:细胞抢救装置通常用于处理心脏手术中排出的红细胞(RBC)。这项研究的目的是评估红细胞功能和体外循环(CPB)期间冲洗过的RBC的质量方面的三种市售细胞保护(CS)设备。 30名接受CPB的患者被随机分配到三个CS设备:C组(Cell Saver 5+; Haemonetics,n = 10),M组(autolog; Medtronic,n = 10)和F组(CATS; Fresenius HemoCare,n = 10)。从水库和输血袋中收集血液样本。分析了水库和洗涤后的RBC的红细胞聚集指数,变形指数(DI)和血细胞比容粘度,2,3-二磷酸甘油酸酯(2,3-DPG),血细胞比容(Hct),血红蛋白(Hb),游离Hb去除率(ΔfHb),葡萄糖(Glu),乳酸(Lac)和血液尿素氮。处理后,与F组相比,C组(P = 0.026)和M(P = 0.032)的红细胞DI相对较高。与M组(P = 0.001)和F(P = 0.001)相比,C组的Δ2,3-DPG较低。 0.001)。 F组提供Hct(P = 0.021; 0.046)和Hb(P = 0.008; 0.013)的最大浓度。另外,C组(P = 0.035)和M组(P = 0.038)的fHb(ΔfHb)去除率更高,与F组有显着差异。然而,基于不同的设计,洗涤的红细胞的功能和不希望的内容物去除效率在一个设备之间到另一个设备上都存在很大差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号