Renal blood flow (RBF) may vary during cardiopulmonary bypass and low flow may cause insufficient blood supply of the kidney triggering renal failure postoperatively. Still, a valid intraoperative method of continuous RBF measurement is not available. A new catheter combining thermodilution and intravascular Doppler was developed, first calibrated in an in vitro model, and the catheter specific constant was determined. Then, application of the device was evaluated in a pilot study in an adult cardiovascular population. The data of the clinical pilot study revealed high correlation between the flow velocities detected by intravascular Doppler and the RBF measured by thermodilution (Pearson's correlation range: 0.78 to 0.97). In conclusion, the RBF can be measured excellently in real time using the new catheter, even under cardiopulmonary bypass.
机译:体外循环期间,肾血流量(RBF)可能会有所不同,低流量可能导致肾脏的血液供应不足,从而引发术后肾衰竭。仍然没有有效的术中连续RBF测量方法。开发了一种将热稀释和血管内多普勒结合起来的新导管,首先在体外模型中进行校准,然后确定导管的比常数。然后,在一项针对成人心血管人群的初步研究中评估了该设备的应用。临床试验研究的数据显示,血管内多普勒检测到的流速与热稀释法测得的RBF之间存在高度相关性(Pearson相关范围:0.78至0.97)。总之,即使在体外循环下,也可以使用新导管实时出色地测量RBF。