首页> 外文期刊>lasers in surgery and medicine >Plasma‐mediated ablation of corneal tissue at 1053 nm using a Nd:YLF oscillator/regenerative amplifier laser
【24h】

Plasma‐mediated ablation of corneal tissue at 1053 nm using a Nd:YLF oscillator/regenerative amplifier laser

机译:Plasma‐mediated ablation of corneal tissue at 1053 nm using a Nd:YLF oscillator/regenerative amplifier laser

获取原文
           

摘要

AbstractPlasma‐mediated ablations were performed on human donor corneas with a short pulsed Nd:YLF laser system at 1053 nm. The pulses were 60 psec in duration at a repetition rate of 1.0 kHz. The laser beam was oriented perpendicular to the cornea surface. The threshold energy densities for ablation of epithelium, Bowman's membrane and stroma were measured. They were 6.1±1.8 J/cm2, 21.0±5.1 J/cm2and 10.4±1.8 J/cm2, respectively. The mean rate of tissue removal at the stromal energy density threshold was about 1 μm per pulse. The walls of the laser excisions were smooth with distortions of less than 1 μm. A new quantitative model of plasma‐mediated ablation is introduced and found to closely predict the observed results. Based on the promising nature of the experimental data further investigations are planned in the use of a mode locked Nd:YLF laser as an alternative to excimer lasers for refractive corneal

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号