Temperature profile of radiofrequency probe application in wrist arthroscopy: Monopolar versus bipolar

机译:腕关节镜中射频探头应用的温度曲线:单极与双极

获取原文
获取原文并翻译 | 示例

摘要

Purpose: The purpose of this study was to investigate the changes in temperature during wrist arthroscopy comparing monopolar and bipolar radiofrequency energy (RFE). Methods: A standard wrist arthroscopy was performed on 14 arms of 7 cadavers without irrigation or with continuous irrigation with 0.9% saline solution and gravity-assisted outflow through an 18-gauge needle. We treated 7 wrists with a bipolar device (VAPR II with 2.3-mm side effect electrodes; DePuy Mitek, Westwood, MA) and 7 wrists with a monopolar device (OPES Ablator for small joints, 45°; Arthrex, Naples, FL). The temperature was recorded simultaneously from 7 predefined anatomic landmarks. Results: We observed an increase in the temperature corresponding to the time of energy application. The highest measured peak temperatures were 52°C (monopolar) and 49.5°C (bipolar) without irrigation. Continuous irrigation led to a significant reduction in the temperature at the site of the energy application. The mean temperature decreased by 7°C for the monopolar system and 5°C for the bipolar system when irrigation was used. For both radiofrequency devices, we found a decrease in the temperature proportional to the distance of the sensors to the radiofrequency probe. Conclusions: Monopolar and bipolar RFE can be safely used in wrist arthroscopy if a continuous irrigation system is applied and the energy impulse does not exceed 5 to 10 seconds. However, it should be used with great care to avoid local heat damage especially at the cartilage. Clinical Relevance: This basic science study was performed to gain data concerning the temperature in wrist arthroscopy and to broaden the knowledge about the risks when using RFE. Furthermore, we sought to control side effects of RFE by finding the best applied form of RFE regarding duration and pulsation (monopolar/bipolar).
机译:目的:本研究的目的是研究手腕关节镜检查期间温度的变化,比较单极和双极射频能量(RFE)。方法:对7具尸体的14臂进行标准腕关节镜检查,不进行冲洗或用0.9%盐溶液连续冲洗并通过18号针头进行重力辅助流出。我们使用双极器械(带2.3毫米副作用电极的VAPR II;马萨诸塞州韦斯特伍德市的DePuy Mitek)治疗了7个手腕,使用单极器械(45°小关节的OPES消融器;佛罗里达州那不勒斯,Arthrex)治疗了7个手腕。从7个预定义的解剖标志同时记录温度。结果:我们观察到温度升高与能量施加时间相对应。无需灌溉,最高实测峰值温度为52°C(单极)和49.5°C(双极)。连续灌溉导致能源应用场所的温度大大降低。使用灌溉时,单极系统的平均温度降低了7°C,双极系统的平均温度降低了5°C。对于这两种射频设备,我们发现温度的降低与传感器到射频探头的距离成正比。结论:如果应用连续冲洗系统且能量脉冲不超过5至10秒,则单极和双极RFE可安全地用于腕关节镜检查。但是,应非常小心地使用它,以免造成局部热损伤,尤其是在软骨处。临床意义:进行这项基础科学研究是为了获得有关腕关节镜检查温度的数据,并扩大使用RFE时的风险知识。此外,我们试图通过找到关于持续时间和脉动(单极/双极)的最佳RFE应用形式来控制RFE的副作用。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号