首页> 外文期刊>Arthroscopy: the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association >Proper insertion angle is essential to prevent intra-articular protrusion of a knotless suture anchor in shoulder rotator cuff repair.
【24h】

Proper insertion angle is essential to prevent intra-articular protrusion of a knotless suture anchor in shoulder rotator cuff repair.

机译:正确的插入角度对于防止肩袖修复中无节缝缝合锚钉的关节内突出至关重要。

获取原文
获取原文并翻译 | 示例
           

摘要

The advent of new arthroscopic devices has led to the development of novel techniques of arthroscopic rotator cuff repair. However, failure to recognize specific technical aspects and improper application of these devices can lead to complications. We report a case of intra-articular protrusion of knotless anchors (PEEK PushLock SP, 4.5 x 18.5 mm; Arthrex, Naples, FL), used in the lateral row of a suture-bridging technique for arthroscopic rotator cuff repair. This case draws attention to the increased length of such devices when compared with traditional suture anchors, the technical aspects of proper device use, the possible patient-related factors such as implant-patient size mismatch, and the importance of additional imaging for the investigation of failure to progress postoperatively.
机译:新的关节镜装置的出现导致了关节镜肩袖修复的新技术的发展。然而,未能认识到特定技术方面以及这些设备的不当使用可能导致复杂化。我们报告一例关节内突出的无节状锚钉(PEEK PushLock SP,4.5 x 18.5 mm; Arthrex,那不勒斯,佛罗里达州),用于关节镜下旋转袖套的缝合桥接技术的侧排。与传统的缝合锚钉相比,这种情况引起了人们的注意,即这种设备的长度增加,正确使用设备的技术方面,可能与患者相关的因素(例如种植体-患者尺寸不匹配)以及额外影像学对于研究牙本质的重要性。术后未能进展。