A hemodynamic evaluation of the Levitronix Pedivas centrifugal pump and Jostra Hl-20 roller pump under pulsatile and nonpulsatile perfusion in an infant CPB model.

机译:婴儿CPB模型中脉搏和非搏动灌注下Levitronix Pedivas离心泵和Jostra Hl-20滚子泵的血流动力学评估。

获取原文
获取原文并翻译 | 示例
           

摘要

The hemodynamic comparison of the Jostra HL-20 and the Levitronix PediVAS blood pumps is the focus this study, where pressure-flow waveforms and hemodynamic energy values are analyzed in the confines of a pediatric cardiopulmonary bypass circuit.The pseudo pediatric patient was perfused with flow rates between 500 and 900 ml/min (100 ml/min increments) under pulsatile and nonpulsatile mode. The Levitronix continuous flow pump utilized a customized controller to engage in pulsatile perfusion with equivalent pulse settings to the Jostra HL-20 roller pump. Hemodynamic measurements and waveforms were recorded at the precannula location, while the mean arterial pressure was maintained at 40 mm Hg for each test. Glycerin water was used as the blood analog circuit perfusate. At each flow rate 24 trials were conducted yielding a total of 120 experiments (n=60 pulsatile and n=60 nonpulsatile).Under nonpulsatile perfusion the Jostra roller pump produced small values for surplus hemodynamic energy (SHE) due to its inherent pulsatility, while the Levitronix produced values of essentially zero for SHE. When switching to pulsatile perfusion, the SHE levels for both the Jostra and Levitronix pump made considerable increases. In comparing the two pumps under pulsatile perfusion, the Levitronix PediVAS produced significantly more surplus and total hemodynamic energy than did the Jostra roller pump each pump flow rate.The study suggests that the Levitronix PediVAS centrifugal pump has the capability of achieving quality pulsatile waveforms and delivering more SHE to the pseudo patient than the Jostra HL-20 roller pump. Further studies are warranted to investigate the Levitronix under bovine blood studies and with various pulsatile settings.
机译:这项研究的重点是Jostra HL-20和Levitronix PediVAS血泵的血流动力学比较,其中在小儿心肺旁路回路的范围内分析了压力-流量波形和血流动力学能量值。在脉动模式和非脉动模式下,速率介于500和900 ml / min之间(增量为100 ml / min)。 Levitronix连续流量泵利用定制的控制器以与Jostra HL-20滚子泵相同的脉冲设置进行脉冲灌注。在插管前位置记录血流动力学测量结果和波形,而每次测试的平均动脉压保持在40 mm Hg。甘油水用作血液模拟回路灌注液。在每种流速下进行了24次试验,总共进行了120次试验(n = 60搏动,n = 60非搏动)。在非搏动灌注下,Jostra滚子泵由于其固有的搏动性而产生的剩余血流动力学能量(SHE)值很小,而Levitronix得出的SHE值基本上为零。当转换为脉动性灌注时,Jostra和Levitronix泵的SHE值均显着增加。在比较两个脉动灌注泵时,Levitronix PediVAS产生的盈余和总血流动力学能量明显高于Jostra滚子泵每个泵的流量。与Jostra HL-20滚轮泵相比,假病人需要更多的SHE。有必要进行进一步的研究来对Levitronix进行牛血研究并在各种搏动环境下进行研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号