Recording the relative blood volume is a standard feature of modern dialysis devices. Three different measurement systems are incorporated in currently available dialysis machines. The Gambro-Hospal group and Nikkiso feature blood volume monitors based on different optical methods. The Fresenius Medical Care machines perform blood volume monitoring with an ultrasonic method. On grounds of clinical evidence suggesting a malfunction of the optical methods in the presence of sodium changes, we compared these three systems. Under the tested conditions, both optical systems show opposite and nonplausible courses of blood volume changes. The ultrasonic system seems to be less susceptible to osmotic changes.
机译:记录相对血容量是现代透析设备的标准功能。当前可用的透析机中并入了三种不同的测量系统。 Gambro-Hospal组和Nikkiso具有基于不同光学方法的血容量监测器。费森尤斯医疗设备使用超声波方法执行血容量监测。基于临床证据表明存在钠变化时光学方法存在故障,我们比较了这三种系统。在测试条件下,两个光学系统均显示出相反的和不合理的血容量变化过程。超声波系统似乎不太容易受到渗透变化的影响。