首页> 外文期刊>Archives of orthopaedic and trauma surgery. >17-year follow-up of the rough-blasted threaded Weill cup in uncemented total hip arthroplasty.
【24h】

17-year follow-up of the rough-blasted threaded Weill cup in uncemented total hip arthroplasty.

机译:在未骨水泥化的全髋关节置换术中对喷砂的螺纹威尔杯进行17年随访。

获取原文
获取原文并翻译 | 示例
           

摘要

INTRODUCTION: For uncemented hip arthroplasty, various cup designs are available. The threaded Weill acetabular component (Weill cup; Zimmer, Winterthur, Switzerland) has been used for more than 20 years, with poor results of the smooth threaded design. Our study was intended to assess the 17-year outcome of the rough-blasted option of the threaded Weill cup. MATERIALS AND METHODS: Between 1987 and 1988, a series of 86 rough-blasted threaded Weill cups were implanted in combination with the CLS Spotorno stem (Zimmer Ltd, Germany) The patients' mean age at the time of surgery was 50 years (range 19-67 years). 67 out of 86 hips (78%) were available for a follow-up at a mean of 17 years (range 16-18 years). Radiographs were available from 55 out of 63 unrevised hips (87%) and analyzed for radiolucency and PE wear. RESULTS: Two out of 86 cups (3%) were revised due to aseptic loosening and another two cups (3%) were awaiting revision for the same reason. Ten patients (10 cups, 12%) were lost to follow-up, and nine patients with nine cups (11%) had deceased without radiographic signs of cup failure. Cup survival with "revision or awaiting revision" as endpoint was 86% (95% CI 75-92%). No deep infections occurred, and no polyethylene insert was exchanged. The Harris hip score was excellent in 37 out of 67 clinically examined hips (55%), good in 18 hips (26%), satisfactory in 5 hips (8%) and moderate or poor in 5 hips (8%) and 2 hips (3%), respectively. CONCLUSION: The rough-blasted threaded Weill cup provides a good long-term performance in cementless total hip arthroplasty. The results compare favourably to the smooth threaded cup design.
机译:简介:对于非骨水泥型髋关节置换术,可以使用各种杯罩设计。带螺纹的Weill髋臼组件(Weill杯; Zimmer,温特图尔,瑞士)已使用20多年,其光滑螺纹设计的效果较差。我们的研究旨在评估螺纹Weill杯的粗加工选项在17年后的结果。材料与方法:在1987年至1988年之间,将一系列86枚粗喷螺纹的威尔杯与CLS Spotorno茎(Zimmer Ltd,德国)一起植入。患者在手术时的平均年龄为50岁(范围19 -67年)。 86例髋关节中有67例(78%)可以接受平均17年(16-18岁)的随访。可以从63例未矫正的髋部中的55例(87%)中获得X线照片,并分析其射线透照度和PE磨损。结果:由于无菌性松动,对86杯中的2杯(3%)进行了修订,出于同样的原因,另外2杯(3%)也在等待修订。失去10例患者(10杯,占12%),而有9杯的9例患者(占11%)死亡,且没有影像学上的杯衰竭迹象。以“修订或待修订”为终点的杯子生存率为86%(95%CI 75-92%)。没有发生深层感染,也没有更换聚乙烯插件。哈里斯(Harris)髋关节得分在67个经过临床检查的髋关节中占37个(55%),在18个髋关节(26%)中良好,在5个髋关节(8%)中令人满意,在5个髋关节(8%)和2个髋关节中度或较差(3%)。结论:经粗喷处理的螺纹威尔杯可在非骨水泥全髋关节置换术中提供良好的长期性能。结果与光滑的螺纹杯设计相比具有优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号