Bioimpedance Measurement of Knee Injuries Using Bipolar Electrode Configuration

机译:使用双极电极配置测量膝关节损伤的生物阻抗

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Currently, there is no suitable solution for the point-of-care diagnosis of knee injuries. A potential portable and low-cost technique for accessing and monitoring knee injuries is bioimpedance measurement. This study validated the feasibility of the bipolar electrode configuration for knee bioimpedance measurement with two electrodes placed on a fixed pair of knee acupuncture locations called Xiyan. Then, the study collected 76 valid samples to investigate the relationship between bioimpedance and knee injuries, among whom 39 patients have unilateral knee injuries, and 37 individuals have healthy knees. The self-contrast results indicated that knee injuries caused a reduction of bioimpedance of the knee by about 5 on average, which was detectable at around 100 kHz (p $approx$ 0.001). Furthermore, the results analyzed by principal component analysis and support vector machines show that the detection sensitivity can reach 87.18 using the leave-one-out cross-validation. We also proposed a low-cost and portable bioimpedance measurement device that meets the needs for measuring knee joint bioimpedance.
机译:目前,膝关节损伤的床旁诊断尚无合适的解决方案。生物阻抗测量是一种潜在的便携式和低成本技术,用于获取和监测膝关节损伤。本研究验证了双极电极配置用于膝关节生物阻抗测量的可行性,将两个电极放置在一对固定的膝关节针灸位置(称为Xiyan)上。然后,该研究收集了76个有效样本,以调查生物阻抗与膝关节损伤之间的关系,其中39例患者为单侧膝关节损伤,37例患者膝关节健康。自对比结果表明,膝关节损伤导致膝关节的生物阻抗平均降低约5%,在100 kHz左右(p $approx$ 0.001)时可检测到。此外,通过主成分分析和支持向量机分析结果表明,采用留一交叉验证的检测灵敏度可以达到87.18%。我们还提出了一种低成本的便携式生物阻抗测量设备,以满足测量膝关节生物阻抗的需要。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号