首页> 外文期刊>Archives of Oral Biology >Effects of enamel and dentine thickness on laser doppler blood-flow signals recorded from the underlying pulp cavity in human teeth in vitro
【24h】

Effects of enamel and dentine thickness on laser doppler blood-flow signals recorded from the underlying pulp cavity in human teeth in vitro

机译:牙釉质和牙本质厚度对人牙齿下牙髓腔记录的激光多普勒血流信号的影响

获取原文
获取原文并翻译 | 示例
       

摘要

Objective: To determine the effect of enamel and dentine thickness on laser Doppler blood-flow (LDF) signals recorded from dental pulp. Design: Observations were made on 18 human premolars that had been extracted from young patients during orthodontic treatment. The apical 2/3 of the root was cut off and the remaining pulp removed. Blood flow signals were recorded from the buccal surface of the crown with a laser Doppler flow metre while dilute blood was pumped at 10 ml/min. through a cannula inserted into the pulp cavity. Recordings were made from the enamel surface and at 0.5 mm steps through the enamel and dentine. Results: The blood flow signal increased significantly as the cavity depth increased and at 2.0 mm, the median flux signal was more than ten times greater than that obtained on the enamel surface. The backscattered light intensity did not change with cavity depth. Conclusion: When recording pulpal blood flow from a human tooth with a laser Doppler flow metre, a substantially better signal-to-noise ratio should be obtained by placing the probe on dentine in the floor of a cavity than on the enamel surface.
机译:目的:确定牙釉质和牙本质厚度对从牙髓记录的激光多普勒血流(LDF)信号的影响。设计:观察在正畸治疗期间从年轻患者中提取的18个人类前磨牙。切掉根的根尖的2/3,并除去剩余的果肉。用激光多普勒流量计从牙冠颊面记录血流信号,同时以10 ml / min的速度泵送稀血。通过插入髓腔的套管插入。从牙釉质表面并以0.5mm的步长通过牙釉质和牙本质进行记录。结果:随着腔深度的增加,血流信号显着增加,并且在2.0 mm处,中值通量信号比牙釉质表面获得的中值通量信号大十倍以上。背向散射光强度不随腔深变化。结论:当用激光多普勒血流计记录人类牙齿的牙髓血流时,将探头放在牙本质底部的牙本质上而不是牙釉质表面上,应该获得实质上更好的信噪比。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号