首页> 外文期刊>Mapan: Journal of Metrology Society of India >Estimation of Uncertainty in the Determination of Serum Electrolytes (Na, K, Ca, Mg) by Flame Atomic Absorption Spectroscopy
【24h】

Estimation of Uncertainty in the Determination of Serum Electrolytes (Na, K, Ca, Mg) by Flame Atomic Absorption Spectroscopy

机译:Estimation of Uncertainty in the Determination of Serum Electrolytes (Na, K, Ca, Mg) by Flame Atomic Absorption Spectroscopy

获取原文
获取原文并翻译 | 示例
           

摘要

From a standardized and validated method to quantify serum electrolytes (Na, K, Ca and Mg) by flame atomic absorption spectroscopy, combined relative uncertainty (CRSU) and the percentage of combined relative standard uncertainty (% CRSU) were determined. For this, analytical grade standards were used and later from a certified reference material (NIST2670a). Previously, the sources of uncertainty used in the measurement using the Ishikawa diagram (cause-effect) were established, including mass concentration, calibration curves, volume, reactive targets, certified reference material, dilution and repeatability. Quantification of the metals was performed using the atomic absorption spectroscopy technique employing a Perkin Elmer model 3100 equipment. The values obtained for the CRSU on average are below 0.030 and for the % CRSU on average they are below 3.1%. These results indicate that the uncertainty values are acceptable since they are below the recommended values for these metals.