Purpose. See-through head-mounted displays (HMDs) can be used to view fluoroscopic imaging during orthopedic surgical procedures. The goals of this study were to determine whether HMDs reduce procedure time, number of fluoroscopic images required, or number of head turns by the surgeon compared with standard monitors. Methods. Sixteen orthopedic surgery residents each performed fluoroscopy-guided drilling of 8 holes for placement of tibial nail distal interlocking screws in an anatomical model, with 4 holes drilled while using HMD and 4 holes drilled while using a standard monitor. Procedure time, number of fluoroscopic images needed, and number of head turns by the resident during the procedure were compared between the 2 modalities. Statistical significance was set at P < .05. Results. Mean (SD) procedure time did not differ significantly between attempts using the standard monitor (55 [37] seconds) vs the HMD (56 [31] seconds) (P = .73). Neither did mean number of fluoroscopic images differ significantly between attempts using the standard monitor vs the HMD (9 [5] images for each) (P = .84). Residents turned their heads significantly more times when using the standard monitor (9 [5] times) vs the HMD (1 [2] times) (P < .001). Conclusions. Head-mounted displays lessened the need for residents to turn their heads away from the surgical field while drilling holes for tibial nail distal interlocking screws in an anatomical model; however, there was no difference in terms of procedure time or number of fluoroscopic images needed using the HMD compared with the standard monitor.
机译:目的。透明头戴式显示器 (HMD) 可用于在骨科手术过程中查看透视成像。本研究的目的是确定与标准监护仪相比,HMD 是否减少了手术时间、所需的透视图像数量或外科医生转头的次数。方法。16 名骨科住院医师每人进行了 8 个孔的透视引导钻孔,用于在解剖模型中放置胫骨钉远端互锁螺钉,其中 4 个孔在使用 HMD 时钻孔,4 个孔在使用标准监视器时钻孔。比较了两种方式的手术时间、所需的透视图像数量和住院医师在手术过程中转头的次数。统计学意义设定为P < .05。结果。使用标准监视器(55 [37] 秒)与 HMD(56 [31] 秒)的尝试之间的平均 (SD) 手术时间没有显着差异 (P = .73)。使用标准监视器与HMD的尝试之间的透视图像平均数量也没有显着差异(每次9 [5]张图像)(P = .84)。使用标准显示器时,居民转头的次数明显更多(9 [5] 次)与 HMD(1 [2] 次)(P < .001)。结论。头戴式显示器减少了住院医师在解剖模型中为胫骨甲远端互锁螺钉钻孔时将头从手术区域转开的需要;然而,与标准监视器相比,使用HMD在手术时间或所需的透视图像数量方面没有差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号