We present an experimental study of a composite voltage-biased superconducting bolometer (VSB). The tested VSB consists of a Ti-film superconducting thermometer (T-c similar to 375 mK) on a Si substrate suspended by NbTi superconducting leads. A resistor attached to the substrate provides calibrated heat input into the bolometer. The current through the bolometer is measured with a superconducting quantum interference device ammeter. Strong negative electrothermal feedback fixes the bolometer temperature at T, and reduces the measured response time from 2.6 s to 13 ms. As predicted, the measured current responsivity of the bolometer is equal to the inverse of the bias voltage. A noise equivalent power of 5 x 10(-17) W/root Hz was measured for a thermal conductance G similar to 4.7 x 10(-10) W/K, which is consistent with the expected thermal noise. Excess noise was observed for bias conditions for which the electrothermal feedback strength was close to maximum. (C) 1998 Optical Society of America. [References: 12]
机译:我们提出了复合电压偏置超导辐射热测量仪(VSB)的实验研究。经过测试的VSB包括一个由NbTi超导引线悬挂的Si衬底上的Ti膜超导温度计(T-c类似于375 mK)。附着在基板上的电阻器将校准后的热量输入到辐射热计中。通过辐射热计的电流是用超导量子干涉仪电流表测量的。强大的负电热反馈将辐射热计的温度固定在T,并将测得的响应时间从2.6 s缩短至13 ms。如预测的那样,测辐射热计的电流响应率等于偏置电压的倒数。对于类似于4.7 x 10(-10)W / K的导热系数G,测得的噪声等效功率为5 x 10(-17)W / root Hz,这与预期的热噪声一致。对于电热反馈强度接近最大值的偏置条件,观察到了过多的噪声。 (C)1998年美国眼镜学会。 [参考:12]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号