首页> 外文期刊>Applied optics >Investigation of the wavelength accuracy of Brewer spectrophotometers
【24h】

Investigation of the wavelength accuracy of Brewer spectrophotometers

机译:Brewer分光光度计的波长精度研究

获取原文
获取原文并翻译 | 示例
           

摘要

The wavelength accuracy of the Brewer spectrophotometer in the 280-360-nm spectral range is improved by a new grating-drive mechanism and a new dispersion function derived from the Brewer geometry. With the new mechanism, the reproducibility of wavelength settings for spectral emission lines is better than 0.3 pm (0.0003 nm) and, with a thermally compensated version, the effect of temperature is less than 0.4 pm K-1. The new dispersion function fits spectral line positions better than the standard function and is less prone to extrapolation error. Applying the new function to data from four new and ten standard drives shows that the new drives perform as well as the best of the standard ones and much better than the majority. (C) 1998 Optical Society of America. [References: 7]
机译:通过新的光栅驱动机制和源自Brewer几何结构的新色散函数,提高了Brewer分光光度计在280-360 nm光谱范围内的波长精度。使用新机制,光谱发射线的波长设置可重复性优于0.3 pm(0.0003 nm),而使用热补偿版本时,温度的影响小于0.4 pm K-1。新的色散函数比标准函数更适合光谱线位置,并且不易出现外推误差。将新功能应用于来自四个新驱动器和十个标准驱动器的数据表明,新驱动器的性能与标准驱动器一样好,并且比大多数驱动器要好得多。 (C)1998年美国眼镜学会。 [参考:7]