Elevated blood drug levels obtained from indwelling silicon catheters during oral cyclosporine A administration.

机译:口服环孢菌素A期间从留置式硅导管获得的血液药物水平升高。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Cyclosporine A (CsA) may be bound to and released from the inner surface of central venous catheters resulting in spuriously elevated blood drug levels. We observed this phenomenon in a boy transplanted because of aplastic anemia up to 9 weeks after CsA had been switched from intravenous to oral administration. In comparison to phlebotomy, simultaneous blood sampling from the double lumen catheter resulted in a greater than 10-fold, or two-fold increase in drug levels, respectively, depending on whether or not the line used for prior CsA infusion was selected. Thus, prolonged binding of CsA to the inner surface of venous catheters should also be considered during oral administration.
机译:环孢菌素A(CsA)可能会与中央静脉导管的内表面结合并从中释放,从而导致血液药物水平虚假升高。我们在CsA从静脉给药改为口服给药后长达9周的再生障碍性贫血移植男孩中观察到这种现象。与静脉切开术相比,双腔导管同时采血导致药物水平分别增加了10倍以上或2倍以上,这取决于是否选择了用于先前CsA输注的管线。因此,口服给药时也应考虑CsA与静脉导管内表面的延长结合。