Comparison of sliced and whole lung sets for the LLNL and JAERI torso phantoms using GE detectors.

机译:切片和整个肺集的比较LLNL并使用通用探测器JAERI躯干幻影。

获取原文
获取原文并翻译 | 示例
           

摘要

The performance characteristics of sliced lung sets were compared to lung sets with activity homogeneously distributed throughout the lung tissue substitute material. The activity estimate from planar sources differs from the estimate from homogeneous sources by a factor of 0.88 to 1.09 depending on the photon energy. This error is small compared to other uncertainties commonly encountered in lung counting, i.e., activity deposition in the lung, detector placement, size difference between individuals, etc. Sliced lung sets could be used instead of homogeneous lung sets to test or provide an interim calibration for a lung counting system.
机译:肺片的性能特征集比较肺集与活动均匀分布在整个肺组织替代材料。从平面来源不同于估计均质来源的0.88倍1.09根据光子能量。一般小于其他的不确定性中遇到肺计数,即活动沉积在肺、探测器放置尺寸个体差异等。集可以使用,而不是均匀的肺集测试或提供一个临时校准肺计数系统。