首页> 外文期刊>Health Physics: Official Journal of the Health Physics Society >Measurement of activity distribution using photostimulable phosphor imaging plates in decommissioned 10 mv medical linear accelerator
【24h】

Measurement of activity distribution using photostimulable phosphor imaging plates in decommissioned 10 mv medical linear accelerator

机译:测量活动分配使用photostimulable磷光成像板退役10 mv医用直线加速器

获取原文
获取原文并翻译 | 示例
           

摘要

Photonuclear reactions generate neutrons in the head of the linear accelerator. Therefore, some parts of the linear accelerator can become activated. Such activated materials must be handled as radioactive waste. The authors attempted to investigate the distribution of induced radioactivity using photostimulable phosphor imaging plates. Autoradiographs were produced from some parts of the linear accelerator (the target, upper jaw, multileaf collimator and shielding). The levels of induced radioactivity were confirmed to be non-uniform within each part from the autoradiographs. The method was a simple and highly sensitive approach to evaluating the relative degree of activation of the linear accelerators, so that appropriate materials management procedures can be carried out.
机译:光致反应产生中子直线加速器。部分线性加速器可以成为激活。处理放射性废物。试图调查的分布使用photostimulable感生放射性磷光成像板。产生的某些部分线性的加速器(目标、上颌multileaf准直器和屏蔽)。放射性物质被证实是不均匀的在每个部分的放射能照像。方法是一个简单和高度敏感的方法评估的相对程度的激活的线性加速器,所以适当的材料管理程序可以进行出去了。