Dose Comparison Using Thermoluminescent Dosimeters During Multislice Computed Tomography With Different Parameters for Simulated Spine Tumor Examination

机译:剂量比较使用热释光的测试仪在多层螺旋计算机断层扫描不同的参数来模拟脊柱肿瘤检查

获取原文
获取原文并翻译 | 示例
           

摘要

This study aims to compare the effect of Philips' Brilliance 64-slice and 256-slice (multislice) computed tomography on effective doses when changing the operating parameters for simulated examinations of patients' spine tumors, including changes in pitch, tube voltage (kV), and effective tube current-time product (mA s). This study considers the possibility of using other probable operating conditions to reduce patients' effective doses. The absorbed doses to organs and skin are measured by taking data from thermoluminescent dosimeters (GR-200 and GR-200F) in relevant positions on the anthropomorphic Rando phantom. We also used an American College of Radiology computed tomography accreditation phantom to experiment with image spatial resolution under various scan conditions in order to achieve results over 5 line pairs per cm, the analytical capability required to meet diagnostic needs. The results show that, in general, when we change the pitch, effective tube current-time product, and tube voltage, the effective doses from 256-slice computed tomography exceed those from 64-slice computed tomography.
机译:本研究旨在比较飞利浦的效果辉煌64片和256片(多层)计算机断层扫描在有效剂量改变操作参数模拟检查病人的脊柱肿瘤,包括音高的变化,管电压(kV)有效管当前时间产品(mA)。研究考虑使用其他的可能性可能的操作条件来减少病人的有效剂量。皮肤测量数据热释光的测试仪(gr - 200和gr - 200 f)在相关位置拟人化兰杜幻影。放射学的计算机断层扫描认证幻影与图像空间实验解决各种扫描条件下实现结果在5线对/厘米,分析能力需要满足诊断的需求。改变音高,有效管当前时间产品,管电压,有效的剂量从256 -计算机断层扫描超过从64 -计算机断层扫描。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号