The goal of this research was to investigate the influence of the matrix on the laser-induced spectroscopy of magnesium. Powdered samples were used and were presented to the measurement as thin distributions on adhesive tape. A wide range of NIST certified reference materials were used as samples. With careful sample preparation and correction for sample surface density on the tape (determined by weighing), reasonable consistency in the Mg signal intensity was obtained regardless of sample matrix. Relative error of ~10% and a precision of 10-20% were obtained for the determination of Mg in several certified samples.
机译:这项研究的目的是研究基质对镁的激光诱导光谱的影响。使用粉末状样品,并将其作为细胶带在胶带上进行测量。各种NIST认证的参考材料都用作样品。通过仔细地准备样品和校正带子上的样品表面密度(通过称重确定),无论样品基质如何,都能获得合理的Mg信号强度一致性。测定几种认证样品中的镁,相对误差约为10%,精度为10-20%。