A direct-injection nebulizer (DIN) has been evaluated for liquid sample introduction in helium microwave-induced plasma mass spectrometry (MIP-MS). A short-term stability test with the direct-injection nebulizer was found to give less than 0.5% RSD for twenty minutes. Isotope abundances for K, Ca, Cr, and Se were determined with less than 10% error on all isotopes. All elements have linear dynamic ranges of two to three orders of magnitude. Detection limits are low to sub-ppb for all elements. Ion lens tuning plays an important role in multielement determinations. Oxide, hydroxide, and doubly charged ions are determined for Cr. As. Sr. Cs, Ba. and Pb. The effect of 10° c methanol and high-dissolved salts is investigated.
机译:在氦微波诱导的等离子体质谱法(MIP-MS)中,已经对直接注入雾化器(DIN)的液体样品引入进行了评估。发现使用直喷雾化器进行的短期稳定性测试在20分钟内得出的RSD小于0.5%。确定所有同位素的K,Ca,Cr和Se同位素丰度均小于10%。所有元素的线性动态范围为2到3个数量级。所有元素的检出限均低于ppb。离子透镜调整在多元素测定中起着重要作用。确定了Cr的氧化物,氢氧根和双电荷离子。如。 Cs爸和铅。研究了10°C甲醇和高溶解盐的影响。