Verification of dose delivery for a prostate sIMRT treatment using a SLIC-EPID.

机译:验证使用SLIC-EPID进行前列腺sIMRT治疗的剂量。

获取原文
获取原文并翻译 | 示例
           

摘要

The current work focuses on the verification of transmitted dose maps, measured using a scanning liquid ionization chamber-electronic portal imaging device (SLIC-EPID) for a typical step-and-shoot prostate IMRT treatment using an anthropomorphic phantom at anterior-posterior (A-P), and several non-zero gantry angles. The dose distributions measured using the SLIC-EPID were then compared with those calculated in the modelled EPID for each segment/subfield and also for the corresponding total fields using a gamma function algorithm with a distance to agreement and dose difference criteria of 2.54mm and 3%, respectively.
机译:当前的工作集中在验证透射剂量图,该剂量图是使用扫描液体电离室-电子门禁成像设备(SLIC-EPID)进行的,用于典型的即插即用前列腺IMRT治疗,其前后使用拟人化体模(AP) ),以及几个非零的龙门角度。然后将使用SLIC-EPID测得的剂量分布与在模型EPID中针对每个段/子场以及相应总场的剂量分布进行比较,并使用伽马函数算法,其一致距离和剂量差标准为2.54mm和3 %, 分别。