Accelerator mass spectrometry measurement of Pu-240/Pu-239 isotope ratios in Novaya Zemlya and Kara Sea sediments

机译:加速器质谱法测量Novaya Zemlya和Kara Sea沉积物中Pu-240 / Pu-239同位素的比率

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Generally low levels of plutonium in environmental samples, often combined with limited sample sizes, necessitate reliable low-level techniques for determination of Pu isotopes. Accelerator mass spectrometry (AMS) has proved to be a powerful method for measuring low-level Pu activity concentrations and Pu isotope ratios. Based on procedural blanks, detection limits for AMS were below 1 fg Pu (equivalent to ca. 2 muBq Pu-139), which can compete with both TIMS, high sensitivity ICP-MS, and certainly alpha-spectrometry, while showing less interference, memory and matrix effects as compared to routine ICP-MS techniques. In addition to low detection limits, the technique offers the advantage of giving information on Pu isotope ratios. Measurements of sediments collected from dumping sites at Novaya Zemlya showed deviation from global fallout Pu-240/(239)pu ratios. (C) 2004 Elsevier Ltd. All rights reserved.
机译:通常,环境样品中的low含量较低,并且常常与有限的样品量结合使用,因此需要可靠的低含量技术来测定Pu同位素。加速器质谱(AMS)已被证明是测量低含量Pu活性浓度和Pu同位素比的有效方法。根据程序空白,AMS的检出限低于1 fg Pu(约等于2 muBq Pu-139),可以与TIMS,高灵敏度ICP-MS和肯定的α光谱法竞争,同时显示较少的干扰,与常规ICP-MS技术相比,具有更好的记忆和基质效应。除了低检测限外,该技术还具有提供有关Pu同位素比的信息的优势。从Novaya Zemlya的倾倒场收集的沉积物的测量结果表明,其偏离了全球沉降物Pu-240 /(239)pu的比率。 (C)2004 Elsevier Ltd.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号