首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Comparison of doses calculated by the Monte Carlo method and measured by LiF TLD in the buildup region for a 60Co photon beam.
【24h】

Comparison of doses calculated by the Monte Carlo method and measured by LiF TLD in the buildup region for a 60Co photon beam.

机译:在60Co光子束的累积区域中,通过蒙特卡罗方法计算并由LiF TLD测量的剂量比较。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

This work studied the percent depth doses of (60)Co photon beams in the buildup region of a plastic phantom by LiF TLD measurements and by Monte Carlo calculations. An agreement within +/-1.5% was found between PDDs measured by TLD and calculated by the Monte Carlo method with the TLD in a plastic phantom. The dose in the plastic phantom was scored in voxels, with thickness scaled by physical and electron density. PDDs calculated by electron density scaling showed a better match with PDD(TLD)(MC); the difference is within +/-1.5% in the buildup region for square and rectangular field sizes.
机译:这项工作通过LiF TLD测量和蒙特卡洛计算研究了塑料体模堆积区域中(60)Co光子束的深度剂量百分比。通过TLD测量并通过蒙特卡洛方法与TLD在塑料模型中计算的PDD之间发现在+/- 1.5%之间的一致性。塑料体模中的剂量以体素计分,厚度通过物理和电子密度进行缩放。通过电子密度定标计算得出的PDD与PDD(TLD)(MC)更好地匹配;对于正方形和矩形场大小,在构建区域中的差异在+/- 1.5%之内。