首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Clinical application of GAFCHROMIC (R) EBT film for in vivo dose measurements of total body irradiation radiotherapy
【24h】

Clinical application of GAFCHROMIC (R) EBT film for in vivo dose measurements of total body irradiation radiotherapy

机译:GAFCHROMIC(R)EBT膜在全身辐射放射治疗体内剂量测量中的临床应用

获取原文
获取原文并翻译 | 示例
           

摘要

The GAFCHROMIC (R) EBT film model is a fairly new film product designed for absorbed dose measurements of high-energy photon beams. In vivo dosimetry for total body irradiation (TBI) remains a challenging task due to the extended source-to-surface distance (SSD), low dose rates, and he use of beam spoilers. EBT film samples were used for dose measurements on an anthropomorphic phantom using a TBI setup. Additionally, in vivo measurements were obtained for two TBI patients. Phantom results verified the suitability of the EBT film for TBI treatment in terms of accuracy, reproducibility, and dose linearity. Doses measured were compared to conventional dosimeter measurements using thermoluminescent dosimeters (TLDs), resulting in an agreement of 4.1 % and 6.7% for the phantom and patient measurements, respectively. Results obtained from the phantom and patients confirm that GAFCHROMIC (R) EBT films are a suitable alternative to TLDs as an in vivo dosimeter in TBI radiotherapy. (c) 2007 Elsevier Ltd. All rights reserved.
机译:GAFCHROMIC(R)EBT胶片型号是一种相当新的胶片产品,旨在用于高能光子束的吸收剂量测量。由于放射源到表面的距离(SSD)延长,剂量率低以及使用光束扰流器,体内辐射剂量(TBI)仍然是一项艰巨的任务。使用TBI装置,将EBT胶片样品用于拟人模型上的剂量测量。另外,获得了两名TBI患者的体内测量值。幻影结果证实了EBT膜在TBI治疗中的准确性,可重复性和剂量线性方面的适用性。将所测量的剂量与使用热发光剂量计(TLD)的常规剂量计进行比较,得出的幻像和患者测量值分别为4.1%和6.7%。从体模和患者获得的结果证实,作为TBI放射疗法的体内剂量计,GAFCHROMIC(R)EBT膜是TLD的合适替代品。 (c)2007 Elsevier Ltd.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号