首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Angular dependence of the wall correction factor for air kerma measurements using cylindrical cavity chambers
【24h】

Angular dependence of the wall correction factor for air kerma measurements using cylindrical cavity chambers

机译:使用圆柱型腔室测量空气比释动能的壁校正因子的角度依赖性

获取原文
获取原文并翻译 | 示例
       

摘要

The dependence of the results of air kerma measurements on the orientation of the cylindrical cavity chamber used for beam standardizations is reexamined. The wall attenuation correction factors for the Co-60 and Cs-136 air kerma measurements were obtained for various irradiation angles by extrapolation of the dependences of experimental data and by Monte Carlo simulations. Applying the correction based on extrapolation does not provide an angle-invariant air kerma rate, as desired. In contrast, Monte Carlo simulations, which take into account various affecting phenomena, produce a wall correction factor that makes the corrected air kerma independent of the incidence angle. (c) 2005 Elsevier Ltd. All rights reserved.
机译:重新检查了空气比释动能测量结果对用于光束标准化的圆柱腔室方向的依赖性。通过外推实验数据的依赖性和通过蒙特卡洛模拟,获得了针对各种辐照角度的Co-60和Cs-136空气比释动能测量的壁衰减校正因子。根据外推应用校正无法提供所需的角度不变空气比释动能比。相反,考虑到各种影响现象的蒙特卡洛模拟产生的壁校正因子使校正后的空气比释动能与入射角无关。 (c)2005 Elsevier Ltd.保留所有权利。