首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Calibration and consistency of results of an ionization-chamber secondary standary measuring system for activity
【24h】

Calibration and consistency of results of an ionization-chamber secondary standary measuring system for activity

机译:电离室二级标准活性测定系统结果的校准和一致性

获取原文
获取原文并翻译 | 示例
           

摘要

Calibration in terms of activity of the ionization-chamber secondary standard measuring systems at the PTB is described. The measurement results of a Centronic IG12/A20, a Vinten ISOCAL IV and a radionuclide calibrator chamber for nuclear medicine applications are discussed, their energy-dependent efficiency curves established and the consistency checked using recently evaluated radionuclide decay data. Criteria for evaluating and transferring calibration factors (or efficiencies) are given.
机译:描述了在PTB的电离室二级标准测量系统的活动方面的校准。讨论了用于核医学应用的Centronic IG12 / A20,Vinten ISOCAL IV和放射性核素校准器腔室的测量结果,建立了它们的能量依赖性效率曲线,并使用最近评估的放射性核素衰变数据检查了一致性。给出了评估和传递校准因子(或效率)的标准。