首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Thermoluminescent characteristics of LiF:Mg, Cu, P and CaSO4:Dy for low dose measurement
【24h】

Thermoluminescent characteristics of LiF:Mg, Cu, P and CaSO4:Dy for low dose measurement

机译:用于低剂量测量的LiF:Mg,Cu,P和CaSO4:Dy的热致发光特性

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Thermoluminescence (TL) characteristics for LiF:Mg, Cu, P, and CaSO4:Dy under the homogeneous field of X-ray beams of diagnostic irradiation and its verification using thermoluminescence dosimetry are presented. The irradiation were performed utilizing a conventional X-ray equipment installed at the Hospital Juarez Norte of Mexico. Different thermoluminescence characteristics of two material were studied, such as batch homogeneity, glow curve, linearity, detection threshold, reproducibility, relative sensitivity and fading. Materials were calibrated in terms of absorbed dose to the standard calibration distance and they were positioned in a generic phantom. The dose analysis, verification and comparison with the measurements obtained by the TLD-100 were performed. Results indicate that the dosimetric peak appears at 202 degrees C and 277.5 degrees C for LiF:Mg, Cu, P and CaSO4:Dy, respectively. TL response as a function of X-ray dose showed a linearity behavior in the very low dose range for all materials. However, the TLD-100 is not accurate for measurements below 4 mGy. CaSO4:Dy is 80% more sensitive than TLD-100 and it show the lowest detection threshold, whereas LiF:Mg, Cu, P is 60% more sensitive than TLD-100. All materials showed very good repeatability. Fading for a period of one month at room temperature showed low fading LiF:Mg, Cu, P, medium and high for TLD-100 and CaSO4:Dy. The results suggest that CaSO4:Dy and LiF:Mg, Cu, P are suitable for measurements at low doses used in radiodiagnostic. (C) 2016 Elsevier Ltd. All rights reserved.
机译:提出了LiF:Mg,Cu,P和CaSO4:Dy在诊断辐射的X射线均匀场下的热致发光(TL)特性,并提出了使用热致发光剂量法进行验证的方法。使用安装在墨西哥华雷斯医院的常规X射线设备进行辐照。研究了两种材料的不同热致发光特性,例如批均一性,辉光曲线,线性,检测阈值,重现性,相对灵敏度和褪色。根据吸收剂量将材料校准到标准校准距离,并将它们放置在通用模型中。进行了剂量分析,验证和与TLD-100获得的测量结果的比较。结果表明,对于LiF:Mg,Cu,P和CaSO4:Dy,剂量峰分别出现在202摄氏度和277.5摄氏度。对于所有材料,TL响应作为X射线剂量的函数在非常低的剂量范围内显示出线性行为。但是,TLD-100对于低于4 mGy的测量并不准确。 CaSO4:Dy的敏感性比TLD-100高80%,并且检测阈值最低,而LiF:Mg,Cu,P的敏感性比TLD-100高60%。所有材料均显示出非常好的重复性。在室温下褪色一个月的时间表明,对于LiD:Mg,Cu,P,低褪色的LiF:TLD-100和CaSO4:Dy中高。结果表明,CaSO4:Dy和LiF:Mg,Cu,P适用于放射诊断中低剂量的测量。 (C)2016 Elsevier Ltd.保留所有权利。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号