首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Determination of full-energy peak efficiency at the center position of a through-hole-type clover detector between 0.05 MeV and 3.2 MeV by source measurements and Monte Carlo simulations
【24h】

Determination of full-energy peak efficiency at the center position of a through-hole-type clover detector between 0.05 MeV and 3.2 MeV by source measurements and Monte Carlo simulations

机译:通过源测量和蒙特卡洛模拟确定通孔型三叶草检测器中心位置在0.05 MeV和3.2 MeV之间的全能峰值效率

获取原文
获取原文并翻译 | 示例
           

摘要

Full-energy peak efficiency at the center position of a through-hole-type clover detector was determined by the measurement of standard sources and by Monte Carlo simulation. The coincidence summing under the large-solid-angle condition was corrected using Monte Carlo calculation based on the specific decay scheme for Ba-133, (EU)-E-152,154, and Co-56. This allowed the peak efficiency to be extended from 0.05 MeV to 3.2 MeV with an approximate uncertainty of 3%. (C) 2014 Elsevier Ltd. All rights reserved.
机译:通孔三叶草检测器中心位置的全能峰值效率是通过标准源的测量和蒙特卡洛模拟确定的。基于Ba-133,(EU)-E-152,154和Co-56的特定衰变方案,使用蒙特卡洛计算对大立体角条件下的重合求和进行了校正。这使峰值效率从0.05 MeV扩展到3.2 MeV,不确定度约为3%。 (C)2014 Elsevier Ltd.保留所有权利。