摘要

The detection efficiencies of tritium samples by using liquid scintillation counter with hydrophilic plastic scintillator (PS) was approximately 48% when the sample of 20 muL was held between 2 PS sheets treated by plasma. The activity and count rates showed a good relationship between 400 Bq to 410 KBq mL~-1. The calculated detection limit of 2 min measurement by the PS was 13 Bq mL~-1 when a confidence was 95%. The plasma method for PS produces no radioactive waste.
机译:当将20μL的样品保持在经过等离子体处理的2个PS片之间时,使用带有亲水性塑料闪烁器(PS)的液体闪烁计数器对counter样品的检测效率约为48%。活性和计数率在400 Bq至410 KBq mL〜-1之间显示出良好的关系。当置信度为95%时,通过PS计算得出的2分钟检测限为13 Bq mL〜-1。 PS的等离子体方法不产生放射性废物。