首页> 外文期刊>Clinical & experimental optometry: journal of the Australian Optometrical Association >Repeatability and reproducibility of corneal thickness using SOCT Copernicus HR
【24h】

Repeatability and reproducibility of corneal thickness using SOCT Copernicus HR

机译:角膜的重复性和再现性厚度使用SOCT哥白尼人力资源

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: The aim of this study is to determine the reliability of corneal thickness measurements derived from SOCT Copernicus HR (Fourier domain OCT). Methods: Thirty healthy eyes of 30 subjects were evaluated. One eye of each patient was chosen randomly. Images were obtained of the central (up to 2.0mm from the corneal apex) and paracentral (2.0 to 4.0mm) cornea. We assessed corneal thickness (central and paracentral) and epithelium thickness. The intra-observer repeatability data were analysed using the intra-class correlation coefficient (ICC) for a range of 95 per cent within-subject standard deviation (SW) and the within-subject coefficient of variation (CW). The level of agreement by Bland-Altman analysis was also represented for the study of the reproducibility between observers and agreement between methods of measurement (automatic versus manual). Results: The mean value of the central corneal thickness (CCT) was 542.4 ± 30.1μm (SD). There was a high intra-observer agreement, finding the best result in the central sector with an intra-class correlation coefficient of 0.99, 95 per cent CI (0.989 to 0.997) and the worst, in the minimum corneal thickness, with an intra-class correlation coefficient of 0.672, 95 per cent CI (0.417 to 0.829). Reproducibility between observers was very high. The best result was found in the central sector thickness obtained both manually and automatically with an intra-class correlation coefficient of 0.990 in both cases and the worst result in the maximum corneal thickness with an intra-class correlation coefficient of 0.827. The agreement between measurement methods was also very high with intra-class correlation coefficient greater than 0.91. On the other hand the repeatability and reproducibility for epithelial measurements was poor. Conclusion: Pachymetric mapping with SOCT Copernicus HR was found to be highly repeatable and reproducible. We found that the device lacks an appropriate ergonomic design as proper focusing of the laser beam onto the cornea for anterior segment scanning required that patients were positioned slightly farther away from the machine head-rest than in the setup for retinal imaging.
机译:目的:本研究的目的是确定的角膜厚度测量的可靠性来自SOCT哥白尼人力资源(傅里叶域中10月)。进行了评估。随机抽取的。中央角膜顶点(2.0毫米)近中心(2.0到4.0毫米)角膜。(中部和近中心)和角膜厚度上皮厚度。重复性数据进行分析使用内部类相关系数(ICC)试标准范围的95%偏差(SW)和受试系数的变异(CW)。Bland-Altman分析也为代表研究之间的重现性观察员和协议的方法测量(自动和手动)。中央角膜厚度的平均值(有条件现金转移支付)为542.4±30.1μm (SD)。intra-observer协议,寻找最佳的结果中央部门的内部类相关系数为0.99,95%可信区间(0.989 - 0.997)和最坏的情况下,最低角膜厚度、内部类相关系数为0.672,95%可信区间(0.417 - 0.829)。观察人士是非常高的。发现在中央部门获得的厚度手动和自动的内部类相关系数为0.990这两种情况下,最大最严重的结果角膜厚度与一个内部类的相关性系数为0.827。测量方法也是非常高的内部类相关系数大于0.91. 再现性上皮测量贫穷。哥白尼人力资源被发现高度可重复的和可再生的。一个适当的工效学设计是正确的聚焦的激光束的角膜前部分扫描要求病人定位略远离吗机器比设置的视网膜头枕成像。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号