PERFORMANCE OF AN INFRARED PHOTOACOUSTIC SINGLE GAS ANALYZER IN MEASURING AMMONIA FROM POULTRY HOUSES

机译:红外光声单气体分析仪测量禽舍氨气的性能

获取原文
获取原文并翻译 | 示例
           

摘要

A single-gas photoacoustic analyzer, Chillgard RT, was evaluated for its performance of measuring ammonia (NH3) concentrations in poultry production under laboratory and field conditions, using a multi-gas photoacoustic analyzer, INNOVA 1412, as a reference method. Calibration gases were used to evaluate the cross-interferences of carbon dioxide (CO2), nitrous oxide (N2O), methane (CH4), and water vapor on NH3 measurements by Chillgard RT. The response times of Chillgard RT and INNOVA were measured in the laboratory using four vessels containing broiler litter. Side-by-side field comparisons between the Chillgard RT and INNOVA were conducted on a laying-hen farm and a broiler farm over two five-month periods. A strong linear relationship existed between the Chillgard RT and INNOVA NH3 readings. The NH3 measurement by the Chillgard RT showed no effect by CO2, N2O, CH4, or water vapor under typical operational conditions in poultry operations. The Chillgard RT demonstrated a faster response to NH3 than the IIVNOVA. Following 120-s measurements, the Chillgard RT and IIVNOVA achieved 99.6% and 96.7% of the expected values, respectively, when the INNOVA was configured at a 30-s sampling interval. The IlVNOVA response time was positively affected by water vapor level in the air samples. The Chillgard RT overestimated NH3 concentrations by 3.24%, 10%, and 12% for laying-hen houses, stored boiler litter, and broiler houses, respectively, as compared to the INNOVA. Thus, performance of the Chillgard RT should be carefully validated before used in field NH3 measurements.
机译:使用多气体光声分析仪INNOVA 1412作为参考方法,评估了单气体光声分析仪Chillgard RT在实验室和野外条件下测量家禽生产中氨(NH3)浓度的性能。校准气体用于评估Chillgard RT在NH3测量中对二氧化碳(CO2),一氧化二氮(N2O),甲烷(CH4)和水蒸气的交叉干扰。 Chillgard RT和INNOVA的响应时间是在实验室中使用装有肉鸡垫料的四个容器进行测量的。在两个五个月的时间里,对蛋鸡场和肉鸡场进行了Chillgard RT和INNOVA的并排田间比较。 Chillgard RT和INNOVA NH3读数之间存在很强的线性关系。在家禽养殖的典型运行条件下,Chillgard RT进行的NH3测量显示不受CO2,N2O,CH4或水蒸气的影响。与IIVNOVA相比,Chillgard RT对NH3的响应更快。经过120秒的测量,当将INNOVA配置为30秒的采样间隔时,Chillgard RT和IIVNOVA分别达到了预期值的99.6%和96.7%。 IlVNOVA响应时间受空气样品中水蒸气含量的正影响。与INNOVA相比,Chillgard RT高估了蛋鸡舍,储存的锅炉垃圾和肉鸡舍的NH3浓度分别为3.24%,10%和12%。因此,在用于野外NH3测量之前,应仔细验证Chillgard RT的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号