Organ dose in cardiac dual?energy computed tomography: a Monte Carlo study

机译:在心脏双器官剂量?断层扫描:蒙特卡罗的研究

获取原文
获取原文并翻译 | 示例
           

摘要

Dual-energy computed tomography (DECT) has appeared as a novel approach with the aim of evaluating artery-related diseases. With the advent of DECT, concerns have been raised about the induction of diseases such as cancer due to high radiation exposure of patients. Therefore, the dose received by patients in DECT should be considered. The parameter most commonly used for patient dosimetry is the effective dose (ED). The purpose of this study is to model and validate a DECT scanner by a developed MCNP Monte Carlo code and to calculate the organ doses, the ED, and the conversion factor (k-factor) used in determining ED in the cardiac imaging protocol. To validate the DECT scanner simulation, a standard dosimetry body phantom was modeled in two radiation modes of single energy CT and DECT. The results of simulated CT dose index (CTDI) were compared with those of ImPACT or measurement data. Then dosimetry phantom was replaced by the male and female ORNL phantoms and the organ doses were calculated. The organ doses were also calculated by ImPACT dose software. In the initial validation stage, the minimum and maximum observed relative differences between results of MNCP simulation and measured were 2.77% and 5.79% for the central CTDI and 1.91% and 5.83% for the averaged peripheral CTDI, respectively. The mean ED of simulation and the ImPACT were 3.23 and 5.55 mSv/100 mAs, and the mean k-factor was 0.016 and 0.032 mSv mGy(-1) cm(-1) in the male and female phantoms, respectively. The k-factor obtained for males is close to the currently used k-factor, but the k-factor for females is almost twice.
机译:双能ct (DECT)作为一个新颖的方法,目的是出现评估artery-related疾病。DECT的出现,已经提出了由于感应的疾病,如癌症高辐射的病人。剂量在DECT应该收到的病人考虑。病人剂量测定法是有效剂量(ED)。本研究的目的是模型和验证DECT扫描仪由发达MCNP蒙特卡罗代码和计算器官剂量,艾德,转换因子(k系数)用于确定在心脏成像协议。DECT扫描仪模拟、标准剂量测定法身体的幻影在两种辐射模式建模CT和DECT单能量。模拟CT剂量指数(CTDI)进行了比较或测量数据的影响。剂量测定法幻影被男性所取代女性ORNL幻影和器官剂量计算。通过影响剂量软件。验证阶段,最小值和最大值观察到相对的结果之间的差异MNCP模拟和测量分别为2.77%和5.79%中央CTDI和1.91%和5.83%分别平均CTDI外围。是3.23和ED的模拟和影响5.55 mSv / 100 mAs,意味着增殖系数为0.016和0.032 mSv mGy(1)厘米(1)男性和女性幻影,分别。男性获得接近当前使用增殖系数,但对女性几乎是增殖系数两次。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号