Accuracy assessment of a novel blood pressure measurement device in a South African adult population: Tensoval duo control.

机译:南非成年人口中新型血压测量设备的准确性评估:Tensoval duo控制。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE: The Tensoval duo control device uses a novel combination of auscultatory and oscillometric technology to measure blood pressure. We assessed the accuracy of this device in a South African adult population according to the International Protocol of the European Society of Hypertension. METHODS: Ethical approval was obtained and 33 participants were recruited from Kimberley Hospital Complex (South Africa). Trained observers took sequential same arm blood pressure measurements from each participant, alternating between mercury sphygmomanometery and the device. RESULTS: The Tensoval duo control passed all the phases of the protocol and achieved a mean difference +/- SD of -3.2 +/- 7.2 and 2.0 +/- 6.0 mmHg for systolic and diastolic pressures, respectively. CONCLUSION: This study confirms the accuracy and recommendation for clinical use of the Tensoval duo control device in adults. Accuracy in special patient groups is yet to be assessed, but this novel technology could be particularly valuable in those with vascular disease.
机译:目的:Tensoval二重奏控制装置结合了听诊和示波技术的新颖组合来测量血压。我们根据《欧洲高血压学会国际协议》评估了该设备在南非成年人口中的准确性。方法:获得伦理学批准,并从南非金伯利医院大楼招募了33名参与者。受过训练的观察者从每个参与者那里进行了相同的手臂血压的连续测量,在水银血压计和设备之间交替进行。结果:Tensoval二重对照通过了方案的所有阶段,收缩压和舒张压的平均差+/- SD分别为-3.2 +/- 7.2和2.0 +/- 6.0 mmHg。结论:本研究证实了Tensoval二重奏控制装置在成人中的临床使用的准确性和建议。特殊患者组的准确性尚待评估,但这项新技术在患有血管疾病的患者中可能特别有价值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号