首页> 外文期刊>Blood Pressure Monitoring >Validation of the A&D UA-1020 upper-arm blood pressure monitor with six different-shaped or different-sized cuffs according to the British Hypertension Society protocol
【24h】

Validation of the A&D UA-1020 upper-arm blood pressure monitor with six different-shaped or different-sized cuffs according to the British Hypertension Society protocol

机译:根据英国高血压学会的协议,对带有六个不同形状或不同尺寸袖带的A&D UA-1020上臂式血压计进行验证

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVE: The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure monitor A&D UA-1020 with six different-shaped or different-sized cuffs according to the British Hypertension Society protocol. METHODS: We recruited 136 individuals for four D-ring cuffs (adult, large, small, and medium) and 114 individuals for two cylindrical cuffs (adult and medium). For each participant, we sequentially measured systolic and diastolic blood pressure using a mercury sphygmomanometer (two observers) and UA-1020 (one supervisor) to obtain three pairs of comparisons. RESULTS: For the four D-ring cuffs, the device achieved grade A. The percentage of blood pressure differences within 5, 10, and 15 mmHg was 64, 88, and 96%, respectively, for systolic blood pressure, and 69, 92, and 99%, respectively, for diastolic blood pressure. The average (±SD) of the device-observers differences was -1.2±7.2 mmHg (P=0.0006) and 1.0±6.0 mmHg (P<0.0001) for systolic and diastolic blood pressure, respectively. For the two cylindrical cuffs, the device also achieved grade A. The percentage of blood pressure differences within 5, 10, and 15 mmHg was 65, 87, and 96%, respectively, for systolic blood pressure, and 64, 91, and 98%, respectively, for diastolic blood pressure. The average of the device-observers differences was -0.4±7.3 mmHg (P=0.29) and 2.0±6.1 mmHg (P<0.0001) for systolic and diastolic blood pressure, respectively. CONCLUSION: The UA-1020 device has passed the requirements of the British Hypertension Society protocol with various sizes of D-ring and cylindrical cuffs, and can be recommended in adults.
机译:目的:本研究旨在根据英国高血压学会的协议,评估带有六个不同形状或不同尺寸袖带的自动示波上臂血压监测仪A&D UA-1020的准确性。方法:我们招募了136个人,分别为四个D形环袖口(成人,大,小和中号)和114个人为两个圆柱状袖带(成人和中号)。对于每位参与者,我们依次使用水银血压计(两名观察员)和UA-1020(一名主管)测量收缩压和舒张压,以获得三对比较。结果:对于四个D形环袖带,该设备达到A级。收缩压分别在5、10和15 mmHg之内的血压差异百分比分别为64%,88%和96%,而69、92%舒张压分别为99%和99%。装置观察者差异的平均(±SD)收缩压和舒张压分别为-1.2±7.2 mmHg(P = 0.0006)和1.0±6.0 mmHg(P <0.0001)。对于两个圆柱形袖带,该设备也达到了A级。收缩压在5、10和15 mmHg之内的血压差异百分比分别为65%,87%和96%,而收缩压分别为64%,91%和98% %分别用于舒张压。对于收缩压和舒张压,装置观察者差异的平均值分别为-0.4±7.3 mmHg(P = 0.29)和2.0±6.1 mmHg(P <0.0001)。结论:UA-1020设备已通过英国高血压协会的要求,具有各种尺寸的D形环和圆柱形袖带,可推荐用于成人。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号