首页> 外文期刊>Blood Pressure Monitoring >Validation of A&D TM-2430 upper-arm blood pressure monitor for ambulatory blood pressure monitoring in children and adolescents, according to the British Hypertension Society protocol.
【24h】

Validation of A&D TM-2430 upper-arm blood pressure monitor for ambulatory blood pressure monitoring in children and adolescents, according to the British Hypertension Society protocol.

机译:根据英国高血压协会的协议,对A&D TM-2430上臂式血压计的验证可用于儿童和青少年的动态血压监测。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The A&D TM-2430 ambulatory blood pressure (BP) monitor has been validated in adults but not in a young population. We sought to validate the device monitoring in children and adolescents, according to the British Hypertension Society (BHS) protocol.The A&D TM-2430 is an automated oscillometric upper-arm device for ambulatory BP monitoring. Nine consecutive measurements were taken in 61 children (mean age, 9.8 years; range, 5-15 years) according to the BHS criteria. Overseen by an independent supervisor, measurements were recorded by two observers blinded from each other's readings and from the device readings.The mean difference ± SD between the observers and device measurements was 0.73 ± 1.64 mmHg for systolic blood pressure (SBP) and -1.23 ± 1.65 mmHg for diastolic blood pressure (DBP), respectively, with an interobserver difference of 4 mmHg. The cumulative percentages of differences within 5, 10, and 15 mmHg were 89, 95, and 98% for SBP and 67, 88, and 98% for DBP. The device achieved a grade A rating for SBP and a B grade for DBP.The A&D TM-2430 upper-arm BP monitor has fulfilled the required BHS standards and can be recommended for measuring ambulatory BP in children and adolescent populations.
机译:A&D TM-2430动态血压(BP)监护仪已在成人中经过验证,但在年轻人中并未得到验证。根据英国高血压协会(BHS)的协议,我们试图验证对儿童和青少年的设备监控.A&D TM-2430是用于动态BP监控的自动示波上臂设备。根据BHS标准,对61名儿童(平均年龄9.8岁;范围5-15岁)进行了九次连续测量。在独立的监督员的监督下,两名观察员对彼此的读数和设备读数不知情,记录了测量值。观察者与设备测量值之间的平均差SD为0.73±1.64 mmHg,收缩压(SBP)为-1.23±舒张压(DBP)分别为1.65 mmHg,观察者之间的差为4 mmHg。在5 mmHg,10 mmHg和15 mmHg之内的累积差异百分比对于SBP分别为89%,95%和98%,对于DBP为67%,88%和98%。该设备的SBP等级为A级,DBP等级为B级。A&D TM-2430上臂式BP监护仪已满足必需的BHS标准,可推荐用于测量儿童和青少年人群的动态BP。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号