Since the advent of the implantable defibrillator over 3 decades ago, defibrillation testing has been used to assess whether the device will detect and treat a ventricular arrhythmia. Early in the development of this therapy, defibrillation testing was required since the energy needed to defibrillate a patient was highly dependent on the system (epicardial patches, coils, etc), and the output of the devices was limited. Current defibrillation systems have more efficacious waveforms and optimized electrodes that provide an increased safety margin. However, a small number of patients implanted with current defibrillators still have inadequate safety margins requiring revision of their systems. These are the patients that one wants to identify with defibrillation testing.
机译:自3年前植入式除颤器的出现以来,除颤测试已被用于评估设备是否会检测和治疗心室心律失常。 在这种疗法的开发早期,需要除颤测试,因为除颤患者所需的能量高度依赖于系统(心外膜斑块,线圈等),并且设备的输出受到限制。 当前的除颤系统具有更有效的波形和优化的电极,可提高安全边缘。 但是,少数植入当前除颤器的患者仍然不足以修改其系统。 这些是人们想要通过除颤测试识别的患者。