Development of an accurate oscillometric blood pressure device for low resource settings.

机译:开发用于低资源设置的精确示波血压设备。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To assess and develop an accurate blood pressure measurement device for use in low resource settings and by untrained staff, according to the World Health Organisation guidelines. METHODS: Ninety-nine adults were recruited to validate the device according to the International Protocol of the European Society of Hypertension. All participants provided written informed consent. Patients with an arrhythmia or unclear Korotkoff sounds were excluded. Nine sequential same-arm measurements were taken from each participant alternating between the test device and mercury sphygmomanometry. Differences between the test device and observers were evaluated according to the criteria of the International Protocol and the Association for the Advancement of Medical Instrumentation. RESULTS: The device failed the first assessment of the oscillometric function and required modification to both the deflation rate and the algorithm to fulfil the International Protocol criteria. It then achieved an acceptable mean difference of -0.7 (4.7) mmHg for systolic and -2.0 (4.6) mmHg for diastolic pressure (oscillometric function) and -1.9 (3.8) mmHg and -0.9 (3.3) mmHg for systolic and diastolic pressures, respectively (auscultatory function). CONCLUSION: The Microlife 3AS1-2 is a semi-automated upper arm device with features consistent with low resource requirements. We successfully developed this device for accurate blood pressure measurement in adults according to the International Protocol, through adjustment of the deflation rate and algorithm. The accuracy and user-friendly design of this low-cost device makes it a highly valuable monitor in clinical practice, particularly in low resource settings and for use by untrained staff.
机译:目的:根据世界卫生组织的指导原则,评估和开发一种精确的血压测量设备,以供资源贫乏地区和未经培训的人员使用。方法:根据欧洲高血压学会国际协议,招募了99名成年人来验证该设备。所有参与者均提供了书面知情同意书。心律不齐或Korotkoff声音不清楚的患者被排除在外。从每个参与者进行九次连续的同臂测量,在测试设备和水银血压计之间交替进行。根据《国际议定书》和医学仪器发展协会的标准评估了测试设备与观察者之间的差异。结果:该设备未通过示波器示波功能的首次评估,因此需要对通气率和算法进行修改以符合国际协议标准。然后,它的收缩压平均值达到-0.7(4.7)mmHg,舒张压(示波函数)达到-2.0(4.6)mmHg,收缩压和舒张压达到-1.9(3.8)mmHg和-0.9(3.3)mmHg,分别(听诊功能)。结论:Microlife 3AS1-2是一种半自动化的上臂设备,具有与低资源需求一致的功能。我们通过调整放气速度和算法,成功开发出了该设备,可以根据国际协议对成年人进行准确的血压测量。这种低成本设备的准确性和用户友好的设计使其成为临床实践中非常有价值的监护仪,尤其是在资源匮乏且未经培训的人员使用时。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号