Differences and effects of medium and large adult cuffs on blood pressure readings in individuals with muscular arms

机译:成人中袖带和大型袖带对有手臂肌肉的人的血压读数的差异和影响

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE: This study analyzed systolic and diastolic blood pressure (SBP and DBP) reading differences in individuals with muscular arms, using medium and large adult cuffs. MATERIALS AND METHODS: Resting blood pressures (BPs) were measured in bodybuilders competing at the Mexican National Bodybuilding and Fitness Championship. The means of two bilateral simultaneous arm BP measurements were obtained using two different cuff sizes, 12 cm (medium adult) and 16 cm (large adult). RESULTS: A total of 193 bodybuilders completed the measurements. With an arm circumference greater than 33 cm, the SBP and DBP average taken with the medium adult cuff was higher than that obtained with the large adult cuff, 8.2±10.6 and 1.6±7.4 mmHg, respectively; however, a significant difference was observed only with the SBP. With the medium adult cuff, 48 of 144 individuals showed SBP at least 140 mmHg, whereas with the large adult cuff there were only 17 of 144 individuals. In those participants with an arm circumference less than 33 cm, the BP was nonsignificantly lower with the large cuff, -4.24±9.2 and -2.24±5.4 mmHg for the SBP and DBP, respectively. CONCLUSION: Incorrect cuffing of a muscular arm leads to significant errors in the measurement of SBP and DBP, similar to that observed in miscuffing of nonmuscular arms.
机译:目的:本研究使用中型和大型成人袖带分析了臂肌个体的收缩压和舒张压(SBP和DBP)读数差异。材料与方法:测量在墨西哥全国健美与健身锦标赛上竞争的健美运动员的静息血压(BPs)。使用两个不同的袖带尺寸(中型成年人)12 cm和中型成年人(成人)16 cm,获得两个双侧同时进行手臂BP测量的方法。结果:总共193个健美运动员完成了测量。当臂围大于33 cm时,中型成人袖带的SBP和DBP平均值高于大成人袖带的SBP和DBP平均值,分别为8.2±10.6和1.6±7.4 mmHg。但是,仅在SBP中观察到显着差异。在成人中袖带中,144个人中有48个人显示SBP至少为140 mmHg,而在成人成人袖带中,144个人中只有17个人。在那些手臂周长小于33 cm的参与者中,大袖带的BP显着降低,SBP和DBP分别为-4.24±9.2和-2.24±5.4 mmHg。结论:不正确的袖口肌肉臂袖口会导致SBP和DBP的测量出现重大误差,这与非肌肉臂袖口袖口的错误类似。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号