Reliability of a noninvasive device to measure systemic hemodynamics in hemodialysis patients.

机译:无创设备在血液透析患者中​​测量全身血液动力学的可靠性。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To evaluate the reliability of a noninvasive hemodynamic monitor in hemodialysis patients. METHODS: We enrolled 15 male patients (mean age 63+/-12 years) on stable chronic hemodialysis. Blood pressure and hemodynamic readings were obtained with the Dynapulse 500 Guardian device (Pulsemetric Inc., San Diego, California, USA), which measures systemic hemodynamics on the basis of oscillometric waveforms obtained through a cuff placed over the brachial artery. Measurements were taken sequentially, in duplicate, before, during and after hemodialysis, in the supine, seated and standing positions on four separate midweek dialysis sessions over a 2-week period. RESULTS: The repeatability of the method was tested using 200 pairs of valid measurements. The average values (+/-SD) were 137+/-22 mmHg for systolic blood pressure, 80+/-13 mmHg for diastolic blood pressure, 76+/-15 bpm for heart rate, 1320+/-268 mmHg/s for dP/dtmax, 2.8+/-0.5 l/min/m for cardiac index, and 1455+/-359 dyn/s/cm for systemic vascular resistance. The mean differences (+/-SD of the difference) between readings were 0.1+/-10.4 mmHg for systolic blood pressure, 0.3+/-6.0 mmHg for diastolic blood pressure, -0.2+/-8.0 bpm for heart rate, 0.2+/-234 mmHg/s for dP/dtmax, 0.03+/-0.26 l/min/m for cardiac index and -10+/-177 for systemic vascular resistance, yielding limits of agreement (95%) of -20 to 20 mmHg for systolic blood pressure, -11 to 12 mmHg for diastolic blood pressure, -16 to 17 bpm for heart rate, -458 to 458 mmHg/s for dP/dtmax, -0.5 to 0.5 l/min/m for cardiac index and -338 to 357 dyn/s/cm for systemic vascular resistance. Other hemodynamic parameters fared similarly, and coefficients of variation were all between 7 and 18%. CONCLUSION: We conclude that the Dynapulse 500 Guardian has adequate reliability indices in hemodialysis patients.
机译:目的:评估血液透析患者无创血流动力学监测仪的可靠性。方法:我们招募了15名男性患者(平均年龄63 +/- 12岁)进行稳定的慢性血液透析。使用Dynapulse 500 Guardian设备(Pulsemetric Inc.,圣地亚哥,加利福尼亚,美国)获得血压和血液动力学读数,该设备根据通过放置在肱动脉上方的袖带获得的示波波形来测量全身血液动力学。在2周的时间内,在四个独立的周中透析阶段中,在仰卧,坐姿和站立姿势中依次进行血液透析之前,期间和之后的测量,一式两份。结果:使用200对有效测量值对方法的重复性进行了测试。收缩压的平均值(+/- SD)为137 +/- 22 mmHg,舒张压的平均值为80 +/- 13 mmHg,心率为76 +/- 15 bpm,1320 +/- 268 mmHg / s对于dP / dtmax,对于心脏指数为2.8 +/- 0.5 l / min / m,对于系统血管阻力为1455 +/- 359 dyn / s / cm。读数之间的平均差异(差异的+/- SD)为:收缩压为0.1 +/- 10.4 mmHg,舒张压为0.3 +/- 6.0 mmHg,心率为-0.2 +/- 8.0 bpm,0.2 + dP / dtmax为--234 mmHg / s,心脏指数为0.03 +/- 0.26 l / min / m,全身血管阻力为-10 +/- 177,一致极限(95%)为-20至20 mmHg收缩压,舒张压为-11至12 mmHg,心律为-16至17 bpm,dP / dtmax为-458至458 mmHg / s,心脏指数为-0.5至0.5 l / min / m和-全身血管阻力为338至357 dyn / s / cm。其他血液动力学参数的表现相似,变异系数都在7%至18%之间。结论:我们得出结论,Dynapulse 500 Guardian在血液透析患者中​​具有足够的可靠性指标。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号