Elective replacement of a 4-year-old leadless pacemaker: New approach for end-of-life management

机译:选修替代一个4岁的无情起搏器:终身管理的新方法

获取原文
获取原文并翻译 | 示例

摘要

We report the first case of new technique of replacement of a Micra TPS, due to battery depletion. A 38-year-old patient was admitted due to battery depletion of a TPS, after 44 months of regular pacemaker functioning. After routine implantation of a new TPS, we use a snare loop inserted in the delivery system to capture the old TPS. We believe this approach a good option not to abandon the depleted device, to avoid possible electrical interference or space occupation in right ventricle. This new approach allows to change the strategy during procedure and does not increase significantly the procedure costs.
机译:我们报告了第一例因电池耗尽而更换Micra TPS的新技术。一名38岁的患者因TPS的电池耗尽而入院,经过44个月的常规起搏器功能。在常规植入新TPS后,我们使用插入传送系统的圈套环捕获旧TPS。我们相信这种方法是一个不放弃耗尽的设备的好选择,以避免可能的电干扰或右心室的空间占用。这种新方法允许在手术过程中改变策略,并且不会显著增加手术成本。