Acceleration of 2D-MR fingerprinting by reducing the number of echoes with increased in-plane resolution: a volunteer study

机译:通过减少面内分辨率增加的回声数量来加速2D-MR指纹识别:志愿者研究

获取原文
获取原文并翻译 | 示例
           

摘要

Objective To compare the absolute values and repeatability of magnetic resonance fingerprinting (MRF) with 3000 and 1500 echoes/slice acquired in 41 s and 20 s (MRF3k and MRF1.5k,respectively).Materials and methods MRF3k and MRF1.5k scans based on fast imaging with steady precession (FISP) were conducted using a 3 T scanner.Inter-scan agreement and intra-scan repeatability were investigated in 41 and 28 subjects,respectively.Region-of-interest (ROI) analysis was conducted on T1 values of MRF3k by two raters,and their agreement was evaluated using intraclass correlation coefficients (ICCs).Between MRF3k and MRF1.5k,differences in T1 and T2 values and inter-measurement correlation coefficients (CCs) were investigated.Intra-measurement repeatability was evaluated using coefficients of variation (CVs).A p value < 0.05 was considered statistically significant.Results The ICCs of ROI measurements were 0.77-0.96.Differences were observed between the two MRF scans,but the CCs of the overall ROIs were 0.99 and 0.97 for the T1 and T2 values,respectively.The mean and median CVs of repeatability were equal to or less than 1.58% and 3.13% in each of the ROIs for T1 and T2,respectively;there were some significant differences between MRF3k and MRF1.5k,but they were small,measuring less than 1%.Discussion Both MRF3k and MRF1.5k had high repeatability,and a strong to very strong correlation was observed,with a trend toward slightly higher values in MRF1.5k.
机译:目的比较磁共振指纹图谱(MRF)在41s和20s(分别为MRF3k和MRF1.5k)采集的3000和1500个回波/切片的绝对值和重复性。材料和方法MRF3k和MRF1。使用3T扫描仪进行基于稳定进动快速成像(FISP)的5k扫描。分别对41名和28名受试者的扫描间一致性和扫描内重复性进行了研究。由两名评分员对MRF3k的T1值进行感兴趣区域(ROI)分析,并使用组内相关系数(ICC)评估其一致性。在MRF3k和MRF1之间。5k,研究T1和T2值的差异以及测量间相关系数(CCs)。使用变异系数(CVs)评估测量内重复性。p值<0.05被认为具有统计学意义。结果ROI测量的ICC为0.77-0.96。两次MRF扫描之间存在差异,但T1和T2值的总体ROI的CCs分别为0.99和0.97。在T1和T2的每个ROI中,重复性的平均和中值CV分别等于或小于1.58%和3.13%;MRF3k和MRF1之间存在一些显著差异。5k,但它们很小,测量值不到1%。讨论MRF3k和MRF1。5k具有很高的重复性,并且观察到一种强到非常强的相关性,MRF1中的值有略高的趋势。5公里。

著录项

  • 来源
  • 作者单位

    Department of Diagnostic Imaging and Nuclear Medicine Kyoto University 54 Shogoin Kawaharacho Sakyoku Kyoto 606-8507 Japan;

    Human Brain Research Center Graduate School of Medicine Kyoto University 54 Shogoin Kawaharacho Sakyoku Kyoto 606-8507 Japan;

    Department of Diagnostic Imaging and Nuclear Medicine Kyoto University 54 Shogoin Kawaharacho Sakyoku Kyoto 606-8507 Japan;

    Department of Diagnostic Imaging and Nuclear Medicine Kyoto University 54 Shogoin Kawaharacho Sakyoku Kyoto 606-8507 Japan;

    Department of Diagnostic Imaging and Nuclear Medicine Kyoto University 54 Shogoin Kawaharacho Sakyoku Kyoto 606-8507 Japan;

    Human Brain Research Center Graduate School of Medicine Kyoto University 54 Shogoin Kawaharacho Sakyoku Kyoto 606-8507 Japan;

    Department of Diagnostic Imaging and Nuclear Medicine Kyoto University 54 Shogoin Kawaharacho Sakyoku Kyoto 606-8507 Japan;

    Magnetic Resonance Siemens Healthcare GmbH Henkestrasse 127 Postfach 32 60 91050 Erlangen Germany;

    Magnetic Resonance Siemens Healthcare GmbH Henkestrasse 127 Postfach 32 60 91050 Erlangen Germany;

    Magnetic Resonance Siemens Healthcare GmbH Henkestrasse 127 Postfach 32 60 91050 Erlangen Germany;

    Department of Diagnostic Imaging and Nuclear Medicine Kyoto University 54 Shogoin Kawaharacho Sakyoku Kyoto 606-8507 Japan;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 磁化学分析法;
  • 关键词

    Magnetic resonance imaging; Brain; Fingerprinting; Test-retest reliability;

    机译:磁共振成像;脑;指纹识别;测试 - 保持可靠性;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号