首页> 外文期刊>Annals of Clinical and Laboratory Science: Official Journal of the Association of Clinical Scientists >Effect of Carbamazepine 10, 11-Epoxide on Serum Carbamazepine Measurement Using a New CMIA Assay: Comparison of Values Obtained by Using PETINIA, CEDIA and Liquid Chromatography Combined with Tandem Mass Spectrometry
【24h】

Effect of Carbamazepine 10, 11-Epoxide on Serum Carbamazepine Measurement Using a New CMIA Assay: Comparison of Values Obtained by Using PETINIA, CEDIA and Liquid Chromatography Combined with Tandem Mass Spectrometry

机译:卡马西平10,11-环氧化合物对血清卡马西平测定的影响,采用新的CMIA分析:PETINIA,CEDIA和液相色谱-串联质谱联用获得的值的比较

获取原文
获取原文并翻译 | 示例
           

摘要

Background. Carbamazepine is a classical anticonvulsant that requires therapeutic drug monitoring. We evaluated the effect of carbamazepine 10, 11 epoxide on a new chemiluminescent immunoassay (CMIA) for application on the Architect i1000SR analyzer. Materials and Methods. Carbamazepine concentrations were measured in 40 specimens collected from patients taking carbamazepine using a PETINIA assay (Vista 1500 analyzer), a CEDIA assay (Cobas c501 analyzer) and the new CMIA assay (Architect i1000 analyzer). In addition, carbamazepine and carbamazepine 10, 11-epoxide concentrations were determined using a reference liquid chromatography combined with a tandem mass spectrometry (LC/MS/MS) reference method in another 15 specimens. These specimens were further analyzed using the PETINIA, CEDIA and CMIA assays. Results and Discussion. A good correlation (regression equation: y = 0.9605x+0.2788, n=40, r=0.98) between values obtained by using the CEDIA assay (x-axis) and the CMIA assay (y-axis) was observed but the PETINIA assay showed significant bias compared to the CMIA assay (regression equation: y=0.8191x+0.3069, n=40, r=0.97). The bias was due to high cross-reactivity of epoxide with the PETINIA assay as revealed by comparing carbamazepine values obtained by LC-MS/MS with these three assays. Conclusion. The new CMIA assay is suitable for therapeutic drug monitoring of carbamazepine because values correlated well with the CEDIA assay values (n=40) as well as LC-MS/MS reference method values in 15 specimens.
机译:背景。卡马西平是一种经典的抗惊厥药,需要监测治疗药物。我们评估了卡马西平10,11环氧化物对新的化学发光免疫分析(CMIA)的影响,该分析方法可用于Architect i1000SR分析仪。材料和方法。使用PETINIA分析(Vista 1500分析仪),CEDIA分析(Cobas c501分析仪)和新的CMIA分析(Architect i1000分析仪)从服用卡马西平的患者收集的40个标本中测量卡马西平浓度。此外,在另外15个样品中,使用参考液相色谱结合串联质谱法(LC / MS / MS)来确定卡马西平和卡马西平10、11-环氧化物的浓度。使用PETINIA,CEDIA和CMIA分析进一步分析这些标本。结果与讨论。观察到使用CEDIA测定(x轴)和CMIA测定(y轴)获得的值之间具有良好的相关性(回归方程:y = 0.9605x + 0.2788,n = 40,r = 0.98),但PETINIA测定与CMIA分析相比有明显的偏差(回归方程:y = 0.8191x + 0.3069,n = 40,r = 0.97)。偏差是由于环氧化物与PETINIA分析的高交叉反应性所致,这是通过将LC-MS / MS获得的卡马西平值与这三种分析方法进行比较所揭示的。结论。新的CMIA测定法适用于卡马西平的治疗性药物监测,因为其值与CEDIA测定值(n = 40)以及15个样品中的LC-MS / MS参考方法值具有很好的相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号