摘要

Background: In pedobarography, clinically meaningful comparison of measurements within or between subjects is limited by data variability and measurement error. This study aims to determine the components of the minimal detectable change (MDC) in impulse across all foot regions and the reliability of these measures. Methods: A convenience sample of foot pressures from 108 visits by normal, healthy subjects aged 2-17 years was studied. Each subject had three pedobarograph measurements taken per foot, with six subjects returning for a second visit for assessment of day-to-day variability. Using a five-region mask, segmental impulses were determined, and from these we obtained the coronal plane pressure index (CPPI). Inter-rater, intra-rater, and day-to-day data were analyzed using intraclass correlation coefficients (ICC) to quantify reliability. Variability of the data was analyzed to quantify the MDC. Results: Inter- and intra-rater reliability was high for all measurements while variability was low, indicating small direct measurement error. Generally, the largest contributing factor to the MDC was day-to-day variability. Step-to-step variability was more dependent on foot segment than age although minor age-related changes were noted. Finally, the high relative variability in the CPPI and the medial mid foot impulse resulted in very high MDCs for these measures.
机译:背景:在儿科学中,受试者内部或之间测量值的临床意义比较受到数据可变性和测量误差的限制。本研究旨在确定所有脚部区域的最小可检测变化(MDC)的组成部分以及这些测量的可靠性。方法:对2-17岁的正常健康受试者108次就诊时足部压力的简易样本进行研究。每名受试者每英尺进行三次足径测量,其中六名受试者第二次回访,以评估日常变化。使用五区面罩,测定节段脉冲,并从中获得冠状面压力指数(CPPI)。使用组内相关系数(ICC)对评分员之间、评分员内部和日常数据进行分析,以量化可靠性。对数据的可变性进行分析,以量化MDC。结果:所有测量的评分员间和评分员内信度均较高,而变异性较低,表明直接测量误差较小。总体而言,MDC的最大影响因素是日常变化。与年龄相比,步履间的变异性更依赖于足节段,尽管有轻微的年龄相关变化。最后,CPPI和足中内侧冲动的高度相对变异性导致这些测量的MDC非常高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号