A novel Sc(III) ion-imprinted polymers (Sc-IIP) sorbent was prepared by using Sc(III) ion-8-hydroxyquinoline complex as the template molecule and methacrylic acid, ethylene glycol dimethacrylate (EGDMA) as the functional monomer and cross-linker, respectively. In the present work, a new method using a microcolumn packed with Sc-ilPs as a solid-phase extraction adsorbent was developed for the selective separation and preconcentration of trace amounts of scandium(III) prior to the determination by inductively coupled plasma mass spectrometry. Various factors influencing the separation and preconcentration of trace scandium, such as pH, sample flow rate and breakthrough volume, elution concentration and volume were investigated in detail. Scandium can be quantitatively retained on the MPs sorbent at pH 4.0-6.5 from sample volume up to 300 mL and then eluted efficiently with 5 mL 3% HNO3 solution, therefore an enrichment factor of 60 was accomplished. The maximum adsorption capacity of the ion-imprinted and non-imprinted sorbent was 12.8 mg g~(-1) and 3.5 mg g~(-1), respectively. Under optimal conditions, the detection limit for scandium was 0.013 ng L~(-1) in water, 0.46 ng g~(-1) in geological sample, and the relative standard deviation was 3.2% (n = 11) at the 0.5 μg L~(-1) . The recoveries of Sc in the spiked environmental water samples ranged from 98.0% to 103% with 100 mL of water sample. The proposed method was validated using certified reference materials and applied to the determination of trace scandium in geological samples and environmental water samples with satisfactory results.
机译:以Sc(III)离子-8-羟基喹啉配合物为模板分子,以甲基丙烯酸,乙二醇二甲基丙烯酸酯(EGDMA)为功能单体,通过交联制备了新型的Sc(III)离子印迹聚合物(Sc-IIP)吸附剂。 -linker。在当前的工作中,开发了一种新的方法,该方法使用装有Sc-ilPs的微柱作为固相萃取吸附剂,用于在电感耦合等离子体质谱法测定之前选择性分离和预富集痕量的III(III)。详细研究了影响痕量scan的分离和预富集的各种因素,例如pH,样品流速和穿透体积,洗脱浓度和体积。 dium可以从最大300 mL样品中定量保留在pH 4.0-6.5的MPs吸附剂上,然后用5 mL 3%HNO3溶液有效洗脱,因此富集系数达到60。离子印迹和非印迹吸附剂的最大吸附容量分别为12.8 mg g〜(-1)和3.5 mg g〜(-1)。在最佳条件下,水中detection的检出限为0.013 ng L〜(-1),地质样品中的detection的检出限为0.46 ng g〜(-1),0.5μg时的相对标准偏差为3.2%(n = 11)。 L〜(-1)加标100mL水样后,加标环境水样中Sc的回收率从98.0%到103%。该方法已通过认证的参考材料验证,并用于地质样品和环境水样品中痕量scan的测定,结果令人满意。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号