Development and validation of NIR model using low-concentration calibration range: rapid analysis of Loniceva japonica solution in ethanol precipitation process

机译:使用低浓度校准范围开发和验证NIR模型:乙醇沉淀过程中忍冬的快速分析

获取原文
获取原文并翻译 | 示例
           

摘要

A strategy for both low-concentration calibration range and large number of sample sets selection in the development of a PLS model is studied. A novel approach based on accuracy profile validated the accuracy and precision of the PLS model. The strategy was applied to the determination of chlorogenic acid content in Lonicera japonica using ethanol precipitation by near-infrared (NIR) transmission spectroscopy. The results found the determination coefficient (R~2), standard errors of calibration and prediction (SEC and SEP) were 0.9648, 71.2 ppm and 74.9 ppm, respectively. The further study showed that PLS model could be used to determine chlorogenic acid content based on accuracy profile, which has a lower limit of quantification (LLOQ) ~1700 ppm. Analytical properties such as accuracy, precision, range and linearity from validation criteria also demonstrated the feasibility of the strategy using a low-concentration calibration set in the PLS model, paving the way for analyses in Chinese Herbal Medicine (CHM) applications.
机译:研究了在PLS模型开发中低浓度校准范围和大量样本集选择的策略。一种基于精度分布图的新颖方法验证了PLS模型的准确性和精度。该策略已应用乙醇通过近红外(NIR)透射光谱分析法测定了忍冬中的绿原酸含量。结果发现,测定系数(R〜2),校准和预测的标准误差(SEC和SEP)分别为0.9648、71.2 ppm和74.9 ppm。进一步的研究表明,PLS模型可用于根据准确度曲线确定绿原酸含量,其定量下限(LLOQ)约为1700 ppm。验证标准的准确性,精确度,范围和线性等分析属性也证明了在PLS模型中使用低浓度校准集进行该策略的可行性,为中草药(CHM)应用的分析铺平了道路。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号