This study introduces a flexible ion chromatographic column-switching technique for the direct analysis of myo-inositol in infant formula samples by pulsed amperometric detection. The system, based on two pumps, two columns and three six-pore valves, allowed for free automatic switching between the switching time window (STW) calibration mode and analysis mode. Chromatographic parameters, including chromatographic columns, mobile phases, elution gradient, column-switching parameters, detector waveform parameters, etc., were optimized systematically for the analysis of myo-inositol in infant formula. In this method, both the pretreatment column and analytical column were arranged to be activated and regenerated during their idle time by optimizing the switching program of the three valves, thus increasing the efficiency of the method. The detection potential, E1, of the quadruple-potential waveform was also optimized and it was set to 0.027 V, which offered both excellent sensitivity and a stable baseline. By applying this optimized method to real samples, the determination of myo-inositol in infant formula was achieved in only 20 minutes, with a linear dynamic range from 0.05 mu g mL(-1) to 120 mu g mL(-1), and a limit of detection (LOD) (S/N = 3) of 2.5 mu g g(-1). The results demonstrate the feasibility of this method for routine analysis of myo-inositol in infant formula with the advantages of automation, simple pretreatment, low cost, good stability, and no use of organic reagents.
机译:这项研究介绍了一种灵活的离子色谱柱切换技术,可通过脉冲安培检测直接分析婴儿配方食品样品中的肌醇。该系统基于两个泵,两个色谱柱和三个六通阀,可以在切换时间窗(STW)校准模式和分析模式之间自由地自动切换。系统优化了色谱参数,包括色谱柱,流动相,洗脱梯度,色谱柱切换参数,检测器波形参数等,以分析婴儿配方奶粉中的肌醇。在该方法中,通过优化三个阀门的切换程序,可将预处理塔和分析塔都安排为在其空闲时间内被激活和再生,从而提高了方法的效率。还优化了四电位波形的检测电位E1并将其设置为0.027 V,这提供了出色的灵敏度和稳定的基线。通过将此优化方法应用于实际样品,婴儿配方奶粉中的肌醇的测定仅需20分钟,线性动态范围为0.05μg mL(-1)至120μgmL(-1),并且检测限(LOD)(S / N = 3)为2.5μgg(-1)。结果表明,该方法具有自动化,预处理简单,成本低,稳定性好,不使用有机试剂的优点,可用于婴儿配方奶粉中肌醇的常规分析。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号