首页> 外文期刊>Analytical methods >A very simple and fast analytical method for atmospheric particulate-bound mercury determination
【24h】

A very simple and fast analytical method for atmospheric particulate-bound mercury determination

机译:一种非常简单而快速的分析方法,用于测定大气中的结合汞

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, we present the results obtained for the determination of particulate-bound mercury (PHg) collected on a glass fiber filter by a combustion-AAS technique using a Direct Mercury Analyzer~R (DMA-80 TRICELL; Milestone Inc., Italy). The accuracy of the method was demonstrated by comparison with the U.S. EPA IO-5 method. Sampling was always performed in duplicate using two identical sampling devices arranged side by side. The limit of quantitation of the proposed method was 0.22 ng, which was in the same order as that observed for the U.S. EPA method (0.23 ng), and corresponds to 5.0 pg m~(-3) for a sampling flow of 30 L min~(-1) and a 24 h sampling period. For paired sampling filters, the precision was <10% for PHg concentrations in the range of 6.5 to 29.3 pg m~(-3). For triplicate filters spiked with 0.3000 and 1.000 ng Hg(II), the recovery was (97 ± 2)% and (85 ± 9)%, respectively. The accuracy was checked by analyzing paired sampling filters by both methods (DMA and U.S. EPA) and did not show any significant difference (p > 0.05).
机译:在本文中,我们介绍了使用直接汞分析仪〜R(DMA-80 TRICELL;意大利Milestone Inc.)通过燃烧AAS技术测定玻璃纤维过滤器上收集的颗粒结合汞(PHg)的测定结果。 )。通过与美国EPA IO-5方法进行比较证明了该方法的准确性。始终使用两个并排布置的相同采样设备一式两份进行采样。拟议方法的定量限为0.22 ng,与美国EPA方法的定量限(0.23 ng)相同,并且在30 L min的采样流量下对应于5.0 pg m〜(-3)。 〜(-1)和24小时的采样时间。对于成对的采样过滤器,PHg浓度在6.5至29.3 pg m〜(-3)范围内时,精度为<10%。对于掺有0.3000和1.000 ng Hg(II)的三重过滤器,回收率分别为(97±2)%和(85±9)%。通过使用两种方法(DMA和美国EPA)分析成对的采样过滤器来检查准确性,但没有显示出任何显着差异(p> 0.05)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号