Self-setup of home respiratory polygraphy for the diagnosis of sleep apnea syndrome: cost-efficiency study

机译:睡眠呼吸暂停综合征诊断家庭呼吸自媒体的自我设置:成本效率研究

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose To evaluate the differences in reliability and costs of home respiratory polygraphy (HRP) when installed by the patient and by a nurse, in order to determine the factors affecting and to consider the possible generalization of self-setup procedure. Several HRP devices have been validated for obstructive sleep apnea (OSA) diagnosis but convenience of a nurse intervention in HRP installation has been scarcely studied. Methods This is a prospective and interventional study. About 301 participants were assigned to 2 groups: self-setup and nurse intervention. Sleep study, questionnaires, and diagnostic procedures were performed following the clinical practice in 2016. Signals were considered lost above 3 min, and success of the test was established according to guidelines. Costs were calculated according to a previous multicenter study. Results Both groups (self-setup and nurse intervention) resulted homogeneous in age, gender, BMI, and final diagnosis of OSA. Signal losses during the test were similar in both groups. Slightly higher percentage of unsuccessful tests were obtained in the self-setup procedure (5.3 vs 2.0%, p = 0.121). The costs were similar (107 vs 105 €) in the self-setup group as compared to the nurse setup group. Conclusions The setup of HRP by either the patient or nurse had similar costs and data acquisition. Both installation procedures of HRP were similar regarding test reliability and costs.
机译:None

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号